Brand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: elite V5, elite V6, elite V8 Product Description: V series Patient Monitor including elite V5, elite V6 and elite V8 which can perform long-time continuous monitoring of multiple physiological parameters. Also, it is capable of storing, displaying, analyzing and controlling measurements, and it will indicate alarms in case of abnormity so that doctors and nurses can deal with them in time. The V series Patient Monitor realize the monitoring of physiological parameters by configuration with different parameter modules which include SpO2 (pulse oxygen saturation, pulse rate and SpO2 plethysmogram) with EDAN SpO2 module or Nellcor SPO2 module, NIBP (systolic pressure, diastolic pressure, mean pressure and pulse rate), TEMP, ECG, RESP(respiration), CO2, IBP, C.O. and AG(anesthetic gas), RM(respiratory mechanics), BIS(bispectral index) and ICG(impedance cardiography). Component: No
Recall
- Recall Number
- Z-1146-2026
- Event Number
- 97879
- Firm
- Edan Diagnostics
- FEI Number
- 3009499478
- Product Code
- MHX
- Status
- Open, Classified
- Root Cause
- Device Design
- Initiated
- October 24, 2025
- Posted
- January 22, 2026
- Address
- 9833 Pacific Heights Blvd, San Diego, CA, 92121-4707
Description
Brand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: elite V5, elite V6, elite V8 Product Description: V series Patient Monitor including elite V5, elite V6 and elite V8 which can perform long-time continuous monitoring of multiple physiological parameters. Also, it is capable of storing, displaying, analyzing and controlling measurements, and it will indicate alarms in case of abnormity so that doctors and nurses can deal with them in time. The V series Patient Monitor realize the monitoring of physiological parameters by configuration with different parameter modules which include SpO2 (pulse oxygen saturation, pulse rate and SpO2 plethysmogram) with EDAN SpO2 module or Nellcor SPO2 module, NIBP (systolic pressure, diastolic pressure, mean pressure and pulse rate), TEMP, ECG, RESP(respiration), CO2, IBP, C.O. and AG(anesthetic gas), RM(respiratory mechanics), BIS(bispectral index) and ICG(impedance cardiography). Component: No
As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential cybersecurity issues with various of the firm device.
On 10/24/2025, the firm emailed notifying customers of a product recall. On 11/06/2025, the firm emailed an "URGENT: MEDICAL DEVICE CORRECTION" Letter to customers informing them of labeling issues involving cybersecurity which arose from a communication from the U.S. Food & Drug Administration. -First issue involves an inappropriate publicly routable IP address that was used as an example in the user manual for demonstrating user-configurable IP settings and Second issue involves an inappropriate publicly IP address that was used as the factory default configuration. Customer are instructed to: 1. Immediately examine their inventory and quarantine product subject to correction. 2. If customers may have further distributed affected product, identify their customers and notify them at once of this product correction. The notification to their customers may be enhanced by including the enclosed URGENT MEDICAL DEVICE CORRECTION. 3. For affected product still in their inventory, place copy of the enclosed Notice Before Use and Corrigenda of User Manual into its package before distribution. 4. For affected product that has been distributed or is in use: (1) Confirm that they have received and acknowledged Corrigenda of User Manual. Print out a copy, attach to the User Manual. (2) Confirm the publicly routable IP address 202.114.4.* is NOT used in the User Maintain in the device. If 202.114.4.* is being used, change to appropriate IP settings of device in the Network Setup according to Instruction Instructions for IP Settings Configuration. 5. Complete the attached Acknowledgement and Receipt Form, sign it, and return a scanned copy to the sender of this notification by email. For questions - Contact: EDAN Instruments, Inc. Monday to Friday 09:00-17:00 (UTC+8) Tel:+86-0755-26898326 EDAN Diagnostics, Inc. Monday to Friday 09:00-17:00 (PST) Tel: 1-(858)-750-3066
Worldwide - U.S. Nationwide distribution in the states of CA, KY, PR, FL, CT, IL, PA, RI, NC, NY, LA, TX, MI and the country of Mexico.
342