VENTANA PATHWAY anti-HER-2/neu (4B5) RABBIT MONOCLONAL PRIMARY ANTIBODY (PATHWAY HER2 (4B5)); Model Number: 790-2991, Lot number: 680957, in vitro diagnostic.
Recall
- Recall Number
- Z-1146-2010
- Event Number
- 53597
- Firm
- Ventana Medical Systems Inc
- FEI Number
- 1000125183
- Product Code
- MVC
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- September 17, 2009
- Posted
- March 16, 2010
- Terminated
- April 22, 2010
- Address
- 1910 E Innovation Park Dr, Tucson, AZ, 85755
Description
VENTANA PATHWAY anti-HER-2/neu (4B5) RABBIT MONOCLONAL PRIMARY ANTIBODY (PATHWAY HER2 (4B5)); Model Number: 790-2991, Lot number: 680957, in vitro diagnostic.
Incorrect result: Light staining inconsistencies with lot number 680957 of the Ventana PATHWAY HER-2/neu (4B5) Rabbit Monoclonal Primary Antibody.
The recall was initiated on 09/17/2009 with Ventana Medical Systems forwarding Urgent Voluntary Medical Device Recall letters to all the affected customers advising them to discontinue the use of lot number 680957 and have their laboratory retest with product from a different lot of PATHWAY HER2 (4B5) any breast cases scored as 0, 1+, or 2+ that were not reflexed to an approved in situ hybridization (ISH) method. Customers were also instructed to contact Ventana Customer Support at 1-800-227-2155 and select 1 for Technical Support to speak to a Customer Support agent who will work with then to replace or credit any product from this specific lot number.
Worldwide distribution: USA, Austria, Denmark, Germany, Hungary, Netherlands, Sweden, United Kingdom, Brazil, Canada, India, Korea, Morocco & Thailand
433 units