FDA Recall
Terminated
ISOLOC 6.5 Software Part Number NT-NW-425-520. Software for radiotherapy treatment.
Recall: Z-1146-2007
·
Initiated April 7, 2006
Recall
- Recall Number
- Z-1146-2007
- Event Number
- 38355
- Firm
- Northwest Medical Physics Equipment Inc
- FEI Number
- 1000126033
- Product Code
- IYE
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- April 7, 2006
- Posted
- August 8, 2007
- Terminated
- April 19, 2011
- Address
- 1910 120th Pl SE, Suite 101, Everett, WA, 98208
Description
ISOLOC 6.5 Software Part Number NT-NW-425-520. Software for radiotherapy treatment.
Reason
This correction was for several software bugs which caused image format problems; incorrect ruler display; software crashes; failure to update pixel spacing; display of unattainable treatment table positions; failure to update gantry angles; wrong shifts for two of three axes of treatment table movement.
Action
Software to fix the problems was written and software users were notified of the availability by e-mail on 04/07/2006. They could download the fixed software from the firm's internet website.
Distribution
Customers in USA, Korea, Spain and UK.
Quantity
141 copies