FDA Recall Open, Classified

Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Containing Coiled Cord cable (part number 0012-00-1801, Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65, 0998-UC-0800-31, 0998-UC-0800-33, 0998-UC-0800-52, 0998-UC-0800-53, 0998-00-0800-55

Recall: Z-1145-2023 · Initiated February 7, 2023

Recall

Recall Number
Z-1145-2023
Event Number
91602
Firm
Datascope Corp.
FEI Number
3001418283
Product Code
DSP
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
February 7, 2023
Posted
March 9, 2023
Address
1300 Macarthur Blvd, Mahwah, NJ, 07430-2052

Description

Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Containing Coiled Cord cable (part number 0012-00-1801, Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65, 0998-UC-0800-31, 0998-UC-0800-33, 0998-UC-0800-52, 0998-UC-0800-53, 0998-00-0800-55

Reason

An unexpected shutdown of the IABP may occur due to a failure of the connection between the Coiled Cord cable and the Cable Assembly backplane to the Coiled Cord cable which provides the communication between the display head and base unit.

Action

The firm sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated 2/7/23 to customers. The customers were instructed to do the following actions: Examine your inventory immediately to determine if you have any Cardiosave Hybrid and/or Rescue IABPs. Prior to use of the Cardiosave IABP, inspect the coiled cable cord to ensure that there is no visible damage. Should you experience an unexpected shutdown of the Cardiosave IABP during therapy, or a present frozen or black screen, utilize another IABP to continue therapy. Until an alternative IABP is located you may attempt to restart the IABP. If the IABP remains non-operational, immediately remove from the patient care environment for further product evaluation. During Installation and Replacement of the Helium Tank per the Cardiosave Operating Instructions, please ensure care is taken not to damage the helium tank or the helium tank yoke while inserting and/or removing the helium tank. Follow the instructions for use when changing the helium tank. If user observes any visual damage to the O-ring installed as part of the quick disconnect fitting, contact your Datascope/Getinge service representative. If your device remains inoperable, please contact your service representative to identify the cause and take the necessary actions required. Complete and sign the attached MEDICAL DEVICE CORRECTION - RESPONSE FORM to acknowledge that you have received and understand this notification. Return the completed form to Datascope/Getinge by e-mailing a scanned copy to [email protected] or by faxing the form to 1-877-660-5841. If you have any questions, please contact your Datascope/Getinge representative or call Datascope/Getinge Technical Support at 1-888-943-8872, options 4, 2, 1, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone). Type of Action by the Company: Datascope/Getinge has developed a hardware correction to address this issue. It is important to note that this i

Distribution

Worldwide Distribution: US (nationwide) and OUS (Foreign) ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BRAZIL, BRUNEI , BULGARIA, CANADA, CHILE, CHINA, CZECH REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, GHANA, HONG KONG, ICELAND, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KUWAIT, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MALAYSIA, MEXICO, MOROCCO, MYANMAR, NETHERLANDS, NIGERIA, NORWAY, OMAN, PAKISTAN, PALESTINE, PHILIPPINES, POLAND, PORTUGAL, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, THAILAND, TRINIDAD AND TOBAGO, TURKEY, TURKMENISTAN, UNITED ARAB EMIRATES, UNITED KINGDOM, VIETNAM, AND YEMEN.

Quantity

4062 Total