FDA Recall Terminated

HOTLINE Blood and Fluid Warmer, Mole l HL-90. Designed for use with the Hotline Fluid Warming Set (L-80, L-70 and L-70N) to warm blood and intravenous (I.V.) fluids and deliver them to the patient's intravenous access site at normothermic temperatures under gravity flow conditions.

Recall: Z-1145-2017 · Initiated October 31, 2016

Recall

Recall Number
Z-1145-2017
Event Number
75557
Firm
Smiths Medical ASD, Inc.
FEI Number
2183502
Product Code
LGZ
Status
Terminated
Root Cause
Labeling mix-ups
Initiated
October 31, 2016
Terminated
November 27, 2017
Address
1265 Grey Fox Rd, Saint Paul, MN, 55112-6929

Description

HOTLINE Blood and Fluid Warmer, Mole l HL-90. Designed for use with the Hotline Fluid Warming Set (L-80, L-70 and L-70N) to warm blood and intravenous (I.V.) fluids and deliver them to the patient's intravenous access site at normothermic temperatures under gravity flow conditions.

Reason

HOTLINE Blood and Fluid Warmers included Greek and Danish language labeling instead of the required Finnish labeling.

Action

Smiths Medical sent an Urgent-Medical Device Field Safety Notice dated October 28, 2016, to all affected customers. The letter described the Reason for Field Corrective Action, Risk to Health, and Instructions to Customers. Advised consignees to notify customers and to review and complete the Recall Confirmation Form and return it to Smith Medical via e-mail to [email protected]. For further questions, please call (651) 633-2556.

Distribution

Internationally to Finland

Quantity

33