FDA Recall Terminated

Starburst MRI Semi-Flex Electrode Device, 25cm/Attached Cable/14Ga/6.4F, Product No. H7877001039091, Catalog No. 700-103909; 25cm/14Ga/6.4F, Product No. H7877001026151, Catalog No. 700-102615; and 25cm/14Ga/6.4F, Product No. H7877001026151, Catalog No. 700-102615

Recall: Z-1143-2015 · Initiated December 19, 2014

Recall

Recall Number
Z-1143-2015
Event Number
70388
FEI Number
3017892510
Product Code
GEI
Status
Terminated
Root Cause
Other
Initiated
December 19, 2014
Posted
February 24, 2015
Terminated
November 27, 2015
Address
Angiodynamics 10, Technical Park Glens Falls, NY, 12804

Description

Starburst MRI Semi-Flex Electrode Device, 25cm/Attached Cable/14Ga/6.4F, Product No. H7877001039091, Catalog No. 700-103909; 25cm/14Ga/6.4F, Product No. H7877001026151, Catalog No. 700-102615; and 25cm/14Ga/6.4F, Product No. H7877001026151, Catalog No. 700-102615

Reason

Printed mark on the trocar intended to indicate a 5cm distance from the distal tip of the needle is actually located 4cm from the distal tip of the needle.

Action

AngioDynamics sent an Urgent Medical Device Recall - Immediate Action Required letter dated December 19, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Actions to be taken: Immediately remove the recall product from your inventory (whether in labs, Central Supply, Shipping and Receiving or ANY other location). Segregate this product in a secure location for return to AngioDynamics, Inc. Immediately forward a copy of this recall notification to all sites to which you have distributed affected product. If affected product is located in your institution, please call AngioDynamics, Inc. Customer Service at 1-800-772-6446 between 8:00a.m. and 7:00p.m. (Monday- Friday: Eastern Standard Time) to obtain a replacement or credit for your returned product. Promptly complete, sign and return the enclosed Reply Verification Tracking Form (even if you do not have any product to return) following the directions on this page and the Reply Verification Form.

Distribution

Worldwide Distribution - US Distribution to the states of AZ, AR, FL, MI, MN, MO, NY, TX, VA, WA and WI., and to the countries of Hungary and Spain.

Quantity

43