FDA Recall Terminated

Product 43 consists of all product under product code: HSB and same usage: Item no: 224001016 M/DN R-F IM NAIL 10MM DIA 224001018 M/DN R-F IM NAIL 10MM DIA 224001020 M/DN R-F IM NAIL 10MM DIA 224001022 M/DN R-F IM NAIL 10MM DIA 224001024 M/DN R-F IM NAIL 10MM DIA 224001026 M/DN R-F IM NAIL 10MM DIA 224001028 M/DN R-F IM NAIL 10MM DIA 224001030 M/DN R-F IM NAIL 10MM DIA 224001032 M/DN R-F IM NAIL 10MM DIA 224001034 M/DN R-F IM NAIL 10MM DIA 224001036 M/DN R-F IM NAIL 10MM DIA 224001038 M/DN R-F IM NAIL 10MM DIA 224001040 M/DN R-F IM NAIL 10MM DIA 224001042 M/DN R-F IM NAIL 10MM DIA 224001114 M/DN R-F IM NAIL 11MM DIA 224001116 M/DN R-F IM NAIL 11MM DIA 224001118 M/DN R-F IM NAIL 11MM DIA 224001120 M/DN R-F IM NAIL 11MM DIA 224001122 M/DN R-F IM NAIL 11MM DIA 224001124 M/DN R-F IM NAIL 11MM DIA 224001126 M/DN R-F IM NAIL 11MM DIA 224001128 M/DN R-F IM NAIL 11MM DIA 224001130 M/DN R-F IM NAIL 11MM DIA 224001132 M/DN R-F IM NAIL 11MM DIA 224001134 M/DN R-F IM NAIL 11MM DIA 224001136 M/DN R-F IM NAIL 11MM DIA 224001138 M/DN R-F IM NAIL 11MM DIA 224001140 M/DN R-F IM NAIL 11MM DIA 224001142 M/DN R-F IM NAIL 11MM DIA 224001144 M/DN R-F IM NAIL 11MM DIA 224001146 M/DN R-F IM NAIL 11MM DIA 224001218 M/DN R-F IM NAIL 12MM DIA 224001220 M/DN R-F IM NAIL 12MM DIA 224001224 M/DN R-F IM NAIL 12MM DIA 224001226 M/DN R-F IM NAIL 12MM DIA 224001228 M/DN R-F IM NAIL 12MM DIA 224001230 M/DN R-F IM NAIL 12MM DIA 224001232 M/DN R-F IM NAIL 12MM DIA 224001234 M/DN R-F IM NAIL 12MM DIA 224001236 M/DN R-F IM NAIL 12MM DIA 224001238 M/DN R-F IM NAIL 12MM DIA 224001240 M/DN R-F IM NAIL 12MM DIA 224001242 M/DN R-F IM NAIL 12MM DIA 224001244 M/DN R-F IM NAIL 12MM DIA 224001248 M/DN R-F IM NAIL 12MM DIA 224001318 M/DN R-F IM NAIL 13MM DIA 224001324 M/DN R-F IM NAIL 13MM DIA 224001326 M/DN R-F IM NAIL 13MM DIA 224001328 M/DN R-F IM NAIL 13MM DIA 224001330 M/DN R-F IM NAIL 13MM DIA 224001332 M/DN R-F IM NAIL 13MM DIA 224001334 M/DN R-F IM NAIL 13MM DIA 224001336 M/DN R-F IM NAIL 13MM DIA 224001338 M/DN R-F IM NAIL 13MM DIA 224001340 M/DN R-F IM NAIL 13MM DIA 224001344 M/DN R-F IM NAIL 13MM DIA 224001346 M/DN R-F IM NAIL 13MM DIA 224001348 M/DN R-F IM NAIL 13MM DIA 224001422 M/DN R-F IM NAIL 14MM DIA 224001436 M/DN R-F IM NAIL 14MM DIA 224001438 M/DN R-F IM NAIL 14MM DIA 224001440 M/DN R-F IM NAIL 14MM DIA 224001442 M/DN R-F IM NAIL 14MM DIA 224001444 M/DN R-F IM NAIL 14MM DIA 224001446 M/DN R-F IM NAIL 14MM DIA 224000922 M/DN R-F IM NAIL 9MM DIA 224000926 M/DN R-F IM NAIL 9MM DIA 224000928 M/DN R-F IM NAIL 9MM DIA 224000930 M/DN R-F IM NAIL 9MM DIA 224000932 M/DN R-F IM NAIL 9MM DIA 224000934 M/DN R-F IM NAIL 9MM DIA 224000936 M/DN R-F IM NAIL 9MM DIA 224000942 M/DN R-F IM NAIL 9MM DIA 224000946 M/DN R-F IM NAIL 9MM DIA 225320010 M/DN TIBIAL I/M NAIL 10MM 225326010 M/DN TIBIAL I/M NAIL 10MM 225328010 M/DN TIBIAL I/M NAIL 10MM 225330010 M/DN TIBIAL I/M NAIL 10MM 225332010 M/DN TIBIAL I/M NAIL 10MM 225334010 M/DN TIBIAL I/M NAIL 10MM 225336010 M/DN TIBIAL I/M NAIL 10MM 225338010 M/DN TIBIAL I/M NAIL 10MM 225344010 M/DN TIBIAL I/M NAIL 10MM 225320011 M/DN TIBIAL I/M NAIL 11MM 225322011 M/DN TIBIAL I/M NAIL 11MM 225324011 M/DN TIBIAL I/M NAIL 11MM 225326011 M/DN TIBIAL I/M NAIL 11MM 225328011 M/DN TIBIAL I/M NAIL 11MM 225330011 M/DN TIBIAL I/M NAIL 11MM 225332011 M/DN TIBIAL I/M NAIL 11MM 225334011 M/DN TIBIAL I/M NAIL 11MM 225336011 M/DN TIBIAL I/M NAIL 11MM 225338011 M/DN TIBIAL I/M NAIL 11MM 225340011 M/DN TIBIAL I/M NAIL 11MM 225342011 M/DN TIBIAL I/M NAIL 11MM 225344011 M/DN TIBIAL I/M NAIL 11MM 225346011 M/DN TIBIAL I/M NAIL 11MM 225328012 M/DN TIBIAL I/M NAIL 12MM 225330012 M/DN TIBIAL I/M NAIL 12MM 225332012 M/DN TIBIAL I/M NAIL 12MM 225334012 M/DN TIBIAL I/M NAIL 12MM 225336012 M/DN TIBIAL I/M NAIL 12MM 225338012 M/DN TIBIAL I/M NAIL 12MM 225340012 M/DN TIBIAL I/M NAIL 12MM 225346012 M/DN TIBIAL I/M NAIL 12MM 225324013 M/DN TIBIAL I/M NAIL 13MM

Recall: Z-1141-2016 · Initiated January 11, 2016

Recall

Recall Number
Z-1141-2016
Event Number
72900
FEI Number
1000132037
Product Code
HSB
Status
Terminated
Root Cause
Packaging
Initiated
January 11, 2016
Posted
March 12, 2016
Terminated
April 13, 2017
Address
Zimmer Manufacturing B.V. Turpeaux Industrial Park Rd 1 km 123.4, Mercedita, PR, 00715

Description

Product 43 consists of all product under product code: HSB and same usage: Item no: 224001016 M/DN R-F IM NAIL 10MM DIA 224001018 M/DN R-F IM NAIL 10MM DIA 224001020 M/DN R-F IM NAIL 10MM DIA 224001022 M/DN R-F IM NAIL 10MM DIA 224001024 M/DN R-F IM NAIL 10MM DIA 224001026 M/DN R-F IM NAIL 10MM DIA 224001028 M/DN R-F IM NAIL 10MM DIA 224001030 M/DN R-F IM NAIL 10MM DIA 224001032 M/DN R-F IM NAIL 10MM DIA 224001034 M/DN R-F IM NAIL 10MM DIA 224001036 M/DN R-F IM NAIL 10MM DIA 224001038 M/DN R-F IM NAIL 10MM DIA 224001040 M/DN R-F IM NAIL 10MM DIA 224001042 M/DN R-F IM NAIL 10MM DIA 224001114 M/DN R-F IM NAIL 11MM DIA 224001116 M/DN R-F IM NAIL 11MM DIA 224001118 M/DN R-F IM NAIL 11MM DIA 224001120 M/DN R-F IM NAIL 11MM DIA 224001122 M/DN R-F IM NAIL 11MM DIA 224001124 M/DN R-F IM NAIL 11MM DIA 224001126 M/DN R-F IM NAIL 11MM DIA 224001128 M/DN R-F IM NAIL 11MM DIA 224001130 M/DN R-F IM NAIL 11MM DIA 224001132 M/DN R-F IM NAIL 11MM DIA 224001134 M/DN R-F IM NAIL 11MM DIA 224001136 M/DN R-F IM NAIL 11MM DIA 224001138 M/DN R-F IM NAIL 11MM DIA 224001140 M/DN R-F IM NAIL 11MM DIA 224001142 M/DN R-F IM NAIL 11MM DIA 224001144 M/DN R-F IM NAIL 11MM DIA 224001146 M/DN R-F IM NAIL 11MM DIA 224001218 M/DN R-F IM NAIL 12MM DIA 224001220 M/DN R-F IM NAIL 12MM DIA 224001224 M/DN R-F IM NAIL 12MM DIA 224001226 M/DN R-F IM NAIL 12MM DIA 224001228 M/DN R-F IM NAIL 12MM DIA 224001230 M/DN R-F IM NAIL 12MM DIA 224001232 M/DN R-F IM NAIL 12MM DIA 224001234 M/DN R-F IM NAIL 12MM DIA 224001236 M/DN R-F IM NAIL 12MM DIA 224001238 M/DN R-F IM NAIL 12MM DIA 224001240 M/DN R-F IM NAIL 12MM DIA 224001242 M/DN R-F IM NAIL 12MM DIA 224001244 M/DN R-F IM NAIL 12MM DIA 224001248 M/DN R-F IM NAIL 12MM DIA 224001318 M/DN R-F IM NAIL 13MM DIA 224001324 M/DN R-F IM NAIL 13MM DIA 224001326 M/DN R-F IM NAIL 13MM DIA 224001328 M/DN R-F IM NAIL 13MM DIA 224001330 M/DN R-F IM NAIL 13MM DIA 224001332 M/DN R-F IM NAIL 13MM DIA 224001334 M/DN R-F IM NAIL 13MM DIA 224001336 M/DN R-F IM NAIL 13MM DIA 224001338 M/DN R-F IM NAIL 13MM DIA 224001340 M/DN R-F IM NAIL 13MM DIA 224001344 M/DN R-F IM NAIL 13MM DIA 224001346 M/DN R-F IM NAIL 13MM DIA 224001348 M/DN R-F IM NAIL 13MM DIA 224001422 M/DN R-F IM NAIL 14MM DIA 224001436 M/DN R-F IM NAIL 14MM DIA 224001438 M/DN R-F IM NAIL 14MM DIA 224001440 M/DN R-F IM NAIL 14MM DIA 224001442 M/DN R-F IM NAIL 14MM DIA 224001444 M/DN R-F IM NAIL 14MM DIA 224001446 M/DN R-F IM NAIL 14MM DIA 224000922 M/DN R-F IM NAIL 9MM DIA 224000926 M/DN R-F IM NAIL 9MM DIA 224000928 M/DN R-F IM NAIL 9MM DIA 224000930 M/DN R-F IM NAIL 9MM DIA 224000932 M/DN R-F IM NAIL 9MM DIA 224000934 M/DN R-F IM NAIL 9MM DIA 224000936 M/DN R-F IM NAIL 9MM DIA 224000942 M/DN R-F IM NAIL 9MM DIA 224000946 M/DN R-F IM NAIL 9MM DIA 225320010 M/DN TIBIAL I/M NAIL 10MM 225326010 M/DN TIBIAL I/M NAIL 10MM 225328010 M/DN TIBIAL I/M NAIL 10MM 225330010 M/DN TIBIAL I/M NAIL 10MM 225332010 M/DN TIBIAL I/M NAIL 10MM 225334010 M/DN TIBIAL I/M NAIL 10MM 225336010 M/DN TIBIAL I/M NAIL 10MM 225338010 M/DN TIBIAL I/M NAIL 10MM 225344010 M/DN TIBIAL I/M NAIL 10MM 225320011 M/DN TIBIAL I/M NAIL 11MM 225322011 M/DN TIBIAL I/M NAIL 11MM 225324011 M/DN TIBIAL I/M NAIL 11MM 225326011 M/DN TIBIAL I/M NAIL 11MM 225328011 M/DN TIBIAL I/M NAIL 11MM 225330011 M/DN TIBIAL I/M NAIL 11MM 225332011 M/DN TIBIAL I/M NAIL 11MM 225334011 M/DN TIBIAL I/M NAIL 11MM 225336011 M/DN TIBIAL I/M NAIL 11MM 225338011 M/DN TIBIAL I/M NAIL 11MM 225340011 M/DN TIBIAL I/M NAIL 11MM 225342011 M/DN TIBIAL I/M NAIL 11MM 225344011 M/DN TIBIAL I/M NAIL 11MM 225346011 M/DN TIBIAL I/M NAIL 11MM 225328012 M/DN TIBIAL I/M NAIL 12MM 225330012 M/DN TIBIAL I/M NAIL 12MM 225332012 M/DN TIBIAL I/M NAIL 12MM 225334012 M/DN TIBIAL I/M NAIL 12MM 225336012 M/DN TIBIAL I/M NAIL 12MM 225338012 M/DN TIBIAL I/M NAIL 12MM 225340012 M/DN TIBIAL I/M NAIL 12MM 225346012 M/DN TIBIAL I/M NAIL 12MM 225324013 M/DN TIBIAL I/M NAIL 13MM

Reason

As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.

Action

Zimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached Certificate of Acknowledgment form to [email protected]. 5. If after reviewing this notification you have further questions or concerns please call Customer Service at 1-800-348-2759, or contact your Zimmer Biomet Sales Representative.

Distribution

US Nationwide Distribution

Quantity

1102