FDA Recall Terminated

HORIBA ABX PENTRA 400 I.S.E. Module, Model Numbers: P400ISE110EN02 and P400ISE110US02, All Serial Numbers, Software Versions 5.0.7 and Below. Used to measure absorbance and ion electrodes.

Recall: Z-1141-2013 · Initiated March 20, 2013

Recall

Recall Number
Z-1141-2013
Event Number
64723
Firm
Horiba Instruments, Inc dba Horiba Medical
FEI Number
3002698356
Product Code
JGS
Status
Terminated
Root Cause
Software design
Initiated
March 20, 2013
Posted
April 18, 2013
Terminated
September 26, 2014
Address
34 Bunsen, Irvine, CA, 92618-4210

Description

HORIBA ABX PENTRA 400 I.S.E. Module, Model Numbers: P400ISE110EN02 and P400ISE110US02, All Serial Numbers, Software Versions 5.0.7 and Below. Used to measure absorbance and ion electrodes.

Reason

HORIBA Medical is informing all ABX PENTRA 400 Analyzer with I.S.E. Module customers that there are no Linearity Limit Flags for the I.S.E. assays currently programmed into the ABX PENTRA 400 Analyzer software. All other assays run on this analyzer do have Linearity Limit Flags to alert the user to values that fall below or above the validated linearity range for each assay.

Action

A recall letter, Field Safety Notice & Acknowledgement mailed to the customers March 20, 2013. The letter included the product information, reason for recall, instructions on what to do with the recalled product, and contact information:"If you have any questions or concerns, please contact your local HORIBA Medical representative. We sincerely apologize for any inconvenience this may cause your laboratory. Please contact your Customer Service Representative at 1-888-903-5001 (option 3) for any additional questions related to this customer letter."

Distribution

Nationwide Distribution.

Quantity

905 units