FDA Recall Terminated

Product 42 consists of all product under product code: HSB and same usage: Item no: 225224010 M/DN FEM IM NAIL 10MM DIA 225228010 M/DN FEM IM NAIL 10MM DIA 225230010 M/DN FEM IM NAIL 10MM DIA 225232010 M/DN FEM IM NAIL 10MM DIA 225234010 M/DN FEM IM NAIL 10MM DIA 225236010 M/DN FEM IM NAIL 10MM DIA 225238010 M/DN FEM IM NAIL 10MM DIA 225240010 M/DN FEM IM NAIL 10MM DIA 225242010 M/DN FEM IM NAIL 10MM DIA 225244010 M/DN FEM IM NAIL 10MM DIA 225246010 M/DN FEM IM NAIL 10MM DIA 225228011 M/DN FEM IM NAIL 11MM DIA 225230011 M/DN FEM IM NAIL 11MM DIA 225232011 M/DN FEM IM NAIL 11MM DIA 225234011 M/DN FEM IM NAIL 11MM DIA 225236011 M/DN FEM IM NAIL 11MM DIA 225238011 M/DN FEM IM NAIL 11MM DIA 225240011 M/DN FEM IM NAIL 11MM DIA 225242011 M/DN FEM IM NAIL 11MM DIA 225244011 M/DN FEM IM NAIL 11MM DIA 225246011 M/DN FEM IM NAIL 11MM DIA 225224012 M/DN FEM IM NAIL 12MM DIA 225232012 M/DN FEM IM NAIL 12MM DIA 225234012 M/DN FEM IM NAIL 12MM DIA 225236012 M/DN FEM IM NAIL 12MM DIA 225238012 M/DN FEM IM NAIL 12MM DIA 225240012 M/DN FEM IM NAIL 12MM DIA 225242012 M/DN FEM IM NAIL 12MM DIA 225244012 M/DN FEM IM NAIL 12MM DIA 225246012 M/DN FEM IM NAIL 12MM DIA 225228013 M/DN FEM IM NAIL 13MM DIA 225230013 M/DN FEM IM NAIL 13MM DIA 225232013 M/DN FEM IM NAIL 13MM DIA 225234013 M/DN FEM IM NAIL 13MM DIA 225236013 M/DN FEM IM NAIL 13MM DIA 225238013 M/DN FEM IM NAIL 13MM DIA 225240013 M/DN FEM IM NAIL 13MM DIA 225242013 M/DN FEM IM NAIL 13MM DIA 225244013 M/DN FEM IM NAIL 13MM DIA 225246013 M/DN FEM IM NAIL 13MM DIA 225236014 M/DN FEM IM NAIL 14MM DIA 225238014 M/DN FEM IM NAIL 14MM DIA 225240014 M/DN FEM IM NAIL 14MM DIA 225242014 M/DN FEM IM NAIL 14MM DIA 225244014 M/DN FEM IM NAIL 14MM DIA 225246014 M/DN FEM IM NAIL 14MM DIA 225234015 M/DN FEM IM NAIL 15MM DIA 225240015 M/DN FEM IM NAIL 15MM DIA 225242015 M/DN FEM IM NAIL 15MM DIA 225244015 M/DN FEM IM NAIL 15MM DIA 225242016 M/DN FEM IM NAIL 16MM DIA 225230008 M/DN FEM IM NAIL 8MM DIA 225232008 M/DN FEM IM NAIL 8MM DIA 225234008 M/DN FEM IM NAIL 8MM DIA 225238008 M/DN FEM IM NAIL 8MM DIA 225228009 M/DN FEM IM NAIL 9MM DIA 225230009 M/DN FEM IM NAIL 9MM DIA 225232009 M/DN FEM IM NAIL 9MM DIA 225234009 M/DN FEM IM NAIL 9MM DIA 225236009 M/DN FEM IM NAIL 9MM DIA 225238009 M/DN FEM IM NAIL 9MM DIA 225240009 M/DN FEM IM NAIL 9MM DIA 225242009 M/DN FEM IM NAIL 9MM DIA 225244009 M/DN FEM IM NAIL 9MM DIA 225418010 M/DN HUMERAL I/M NAIL 10M 225419510 M/DN HUMERAL I/M NAIL 10M 225421010 M/DN HUMERAL I/M NAIL 10M 225422510 M/DN HUMERAL I/M NAIL 10M 225424010 M/DN HUMERAL I/M NAIL 10M 225425510 M/DN HUMERAL I/M NAIL 10M 225427010 M/DN HUMERAL I/M NAIL 10M 225428510 M/DN HUMERAL I/M NAIL 10M 225430010 M/DN HUMERAL I/M NAIL 10M 225419511 M/DN HUMERAL I/M NAIL 11M 225421011 M/DN HUMERAL I/M NAIL 11M 225422511 M/DN HUMERAL I/M NAIL 11M 225424011 M/DN HUMERAL I/M NAIL 11M 225425511 M/DN HUMERAL I/M NAIL 11M 225427011 M/DN HUMERAL I/M NAIL 11M 225421012 M/DN HUMERAL I/M NAIL 12M 225422512 M/DN HUMERAL I/M NAIL 12M 225424012 M/DN HUMERAL I/M NAIL 12M 225425512 M/DN HUMERAL I/M NAIL 12M 225427012 M/DN HUMERAL I/M NAIL 12M 225428512 M/DN HUMERAL I/M NAIL 12M 225430012 M/DN HUMERAL I/M NAIL 12M 225418013 M/DN HUMERAL I/M NAIL 13M 225424013 M/DN HUMERAL I/M NAIL 13M 225427013 M/DN HUMERAL I/M NAIL 13M 225418006 M/DN HUMERAL I/M NAIL 6MM 225419506 M/DN HUMERAL I/M NAIL 6MM 225421006 M/DN HUMERAL I/M NAIL 6MM 225424006 M/DN HUMERAL I/M NAIL 6MM 225427006 M/DN HUMERAL I/M NAIL 6MM 225428506 M/DN HUMERAL I/M NAIL 6MM 225430006 M/DN HUMERAL I/M NAIL 6MM 225418007 M/DN HUMERAL I/M NAIL 7MM 225419507 M/DN HUMERAL I/M NAIL 7MM 225421007 M/DN HUMERAL I/M NAIL 7MM 225422507 M/DN HUMERAL I/M NAIL 7MM 225424007 M/DN HUMERAL I/M NAIL 7MM 225425507 M/DN HUMERAL I/M NAIL 7MM 225427007 M/DN HUMERAL I/M NAIL 7MM 225428507 M/DN HUMERAL I/M NAIL 7MM 225430007 M/DN HUMERAL I/M NAIL 7MM 225418008 M/DN HUMERAL I/M NAIL 8MM 225

Recall: Z-1140-2016 · Initiated January 11, 2016

Recall

Recall Number
Z-1140-2016
Event Number
72900
FEI Number
1000132037
Product Code
HSB
Status
Terminated
Root Cause
Packaging
Initiated
January 11, 2016
Posted
March 12, 2016
Terminated
April 13, 2017
Address
Zimmer Manufacturing B.V. Turpeaux Industrial Park Rd 1 km 123.4, Mercedita, PR, 00715

Description

Product 42 consists of all product under product code: HSB and same usage: Item no: 225224010 M/DN FEM IM NAIL 10MM DIA 225228010 M/DN FEM IM NAIL 10MM DIA 225230010 M/DN FEM IM NAIL 10MM DIA 225232010 M/DN FEM IM NAIL 10MM DIA 225234010 M/DN FEM IM NAIL 10MM DIA 225236010 M/DN FEM IM NAIL 10MM DIA 225238010 M/DN FEM IM NAIL 10MM DIA 225240010 M/DN FEM IM NAIL 10MM DIA 225242010 M/DN FEM IM NAIL 10MM DIA 225244010 M/DN FEM IM NAIL 10MM DIA 225246010 M/DN FEM IM NAIL 10MM DIA 225228011 M/DN FEM IM NAIL 11MM DIA 225230011 M/DN FEM IM NAIL 11MM DIA 225232011 M/DN FEM IM NAIL 11MM DIA 225234011 M/DN FEM IM NAIL 11MM DIA 225236011 M/DN FEM IM NAIL 11MM DIA 225238011 M/DN FEM IM NAIL 11MM DIA 225240011 M/DN FEM IM NAIL 11MM DIA 225242011 M/DN FEM IM NAIL 11MM DIA 225244011 M/DN FEM IM NAIL 11MM DIA 225246011 M/DN FEM IM NAIL 11MM DIA 225224012 M/DN FEM IM NAIL 12MM DIA 225232012 M/DN FEM IM NAIL 12MM DIA 225234012 M/DN FEM IM NAIL 12MM DIA 225236012 M/DN FEM IM NAIL 12MM DIA 225238012 M/DN FEM IM NAIL 12MM DIA 225240012 M/DN FEM IM NAIL 12MM DIA 225242012 M/DN FEM IM NAIL 12MM DIA 225244012 M/DN FEM IM NAIL 12MM DIA 225246012 M/DN FEM IM NAIL 12MM DIA 225228013 M/DN FEM IM NAIL 13MM DIA 225230013 M/DN FEM IM NAIL 13MM DIA 225232013 M/DN FEM IM NAIL 13MM DIA 225234013 M/DN FEM IM NAIL 13MM DIA 225236013 M/DN FEM IM NAIL 13MM DIA 225238013 M/DN FEM IM NAIL 13MM DIA 225240013 M/DN FEM IM NAIL 13MM DIA 225242013 M/DN FEM IM NAIL 13MM DIA 225244013 M/DN FEM IM NAIL 13MM DIA 225246013 M/DN FEM IM NAIL 13MM DIA 225236014 M/DN FEM IM NAIL 14MM DIA 225238014 M/DN FEM IM NAIL 14MM DIA 225240014 M/DN FEM IM NAIL 14MM DIA 225242014 M/DN FEM IM NAIL 14MM DIA 225244014 M/DN FEM IM NAIL 14MM DIA 225246014 M/DN FEM IM NAIL 14MM DIA 225234015 M/DN FEM IM NAIL 15MM DIA 225240015 M/DN FEM IM NAIL 15MM DIA 225242015 M/DN FEM IM NAIL 15MM DIA 225244015 M/DN FEM IM NAIL 15MM DIA 225242016 M/DN FEM IM NAIL 16MM DIA 225230008 M/DN FEM IM NAIL 8MM DIA 225232008 M/DN FEM IM NAIL 8MM DIA 225234008 M/DN FEM IM NAIL 8MM DIA 225238008 M/DN FEM IM NAIL 8MM DIA 225228009 M/DN FEM IM NAIL 9MM DIA 225230009 M/DN FEM IM NAIL 9MM DIA 225232009 M/DN FEM IM NAIL 9MM DIA 225234009 M/DN FEM IM NAIL 9MM DIA 225236009 M/DN FEM IM NAIL 9MM DIA 225238009 M/DN FEM IM NAIL 9MM DIA 225240009 M/DN FEM IM NAIL 9MM DIA 225242009 M/DN FEM IM NAIL 9MM DIA 225244009 M/DN FEM IM NAIL 9MM DIA 225418010 M/DN HUMERAL I/M NAIL 10M 225419510 M/DN HUMERAL I/M NAIL 10M 225421010 M/DN HUMERAL I/M NAIL 10M 225422510 M/DN HUMERAL I/M NAIL 10M 225424010 M/DN HUMERAL I/M NAIL 10M 225425510 M/DN HUMERAL I/M NAIL 10M 225427010 M/DN HUMERAL I/M NAIL 10M 225428510 M/DN HUMERAL I/M NAIL 10M 225430010 M/DN HUMERAL I/M NAIL 10M 225419511 M/DN HUMERAL I/M NAIL 11M 225421011 M/DN HUMERAL I/M NAIL 11M 225422511 M/DN HUMERAL I/M NAIL 11M 225424011 M/DN HUMERAL I/M NAIL 11M 225425511 M/DN HUMERAL I/M NAIL 11M 225427011 M/DN HUMERAL I/M NAIL 11M 225421012 M/DN HUMERAL I/M NAIL 12M 225422512 M/DN HUMERAL I/M NAIL 12M 225424012 M/DN HUMERAL I/M NAIL 12M 225425512 M/DN HUMERAL I/M NAIL 12M 225427012 M/DN HUMERAL I/M NAIL 12M 225428512 M/DN HUMERAL I/M NAIL 12M 225430012 M/DN HUMERAL I/M NAIL 12M 225418013 M/DN HUMERAL I/M NAIL 13M 225424013 M/DN HUMERAL I/M NAIL 13M 225427013 M/DN HUMERAL I/M NAIL 13M 225418006 M/DN HUMERAL I/M NAIL 6MM 225419506 M/DN HUMERAL I/M NAIL 6MM 225421006 M/DN HUMERAL I/M NAIL 6MM 225424006 M/DN HUMERAL I/M NAIL 6MM 225427006 M/DN HUMERAL I/M NAIL 6MM 225428506 M/DN HUMERAL I/M NAIL 6MM 225430006 M/DN HUMERAL I/M NAIL 6MM 225418007 M/DN HUMERAL I/M NAIL 7MM 225419507 M/DN HUMERAL I/M NAIL 7MM 225421007 M/DN HUMERAL I/M NAIL 7MM 225422507 M/DN HUMERAL I/M NAIL 7MM 225424007 M/DN HUMERAL I/M NAIL 7MM 225425507 M/DN HUMERAL I/M NAIL 7MM 225427007 M/DN HUMERAL I/M NAIL 7MM 225428507 M/DN HUMERAL I/M NAIL 7MM 225430007 M/DN HUMERAL I/M NAIL 7MM 225418008 M/DN HUMERAL I/M NAIL 8MM 225

Reason

As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.

Action

Zimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached Certificate of Acknowledgment form to [email protected]. 5. If after reviewing this notification you have further questions or concerns please call Customer Service at 1-800-348-2759, or contact your Zimmer Biomet Sales Representative.

Distribution

US Nationwide Distribution

Quantity

1622