B-K Medical Intraoperative Transducer Type 8814, diagnostic ultrasound transducer, B-K Medical Systems, Inc., Wilmington, MA 01887
Recall: Z-1140-2008
·
Initiated January 24, 2008
Recall
- Recall Number
- Z-1140-2008
- Event Number
- 46394
- Firm
- B & K Medical Systems, Inc.
- FEI Number
- 1000305740
- Product Code
- IYN
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- January 24, 2008
- Posted
- June 10, 2008
- Terminated
- June 23, 2008
- Address
- 250 Andover St, Wilmington, MA, 01887-1048
Description
B-K Medical Intraoperative Transducer Type 8814, diagnostic ultrasound transducer, B-K Medical Systems, Inc., Wilmington, MA 01887
Reason
Crack: Incomplete glue joint in the device housing may affect electrical safety or sterilization.
Action
B-K Medical issued letter notification on 1/23/08 to accounts via Fed Ex, advising them not to use the unit until a representative can inspect and exchange, if unit is found defective. Instructions are also provided to continue use of the device with a sterile cover until site visit.
Distribution
USA Distribution: MA, MI, PA, TX, and OR.
Quantity
7