FDA Recall Terminated

IMAGEnet 6 v1.53- IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, storage and management of digital images, patient data, diagnostic data and clinical information from Topcon devices without controlling or altering the functions and parameters of any medical devices or through computerized networks.

Recall: Z-1138-2019 · Initiated March 6, 2019

Recall

Recall Number
Z-1138-2019
Event Number
82322
Firm
Topcon Medical Systems, Inc.
FEI Number
2242863
Product Code
NFJ
Status
Terminated
Root Cause
Labeling design
Initiated
March 6, 2019
Posted
April 12, 2019
Terminated
July 17, 2019
Address
111 Bauer Dr, Oakland, NJ, 07436-3123

Description

IMAGEnet 6 v1.53- IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, storage and management of digital images, patient data, diagnostic data and clinical information from Topcon devices without controlling or altering the functions and parameters of any medical devices or through computerized networks.

Reason

When a patient is selected in the patient data and another patient is selected prior to the display of the primary patient's thumbnails is completed, then the primary patient's thumbnails are displayed with secondary patient's identification information such as name, patient's ID, etc. on the monitor screen of the PC used.

Action

Topcon issued Device Correction Notice on 3/22/19 providing problem- there exists a situation in which the displayed patient name and the thumbnail image do not match and action to take.Topcon has corrected this issue in IMAGEnet 6 version 2.0.3 and will be providing a remote upgrade free of charge. Contact customer support team to schedule the upgrade at 866.922.6278 option 1 or by email at [email protected]

Distribution

Nationwide

Quantity

450 units