MyLab Ultrasound System; Models 6100, 6150, 6200, 6250; Probes: BL433, C 18, CA541, EC1123, EC123, IH 618, IOT332, IOT342, L 415, LA332E, LA435, LA523, LA533, SI2C41, SL2325, TRT33
Recall
- Recall Number
- Z-1138-2018
- Event Number
- 79547
- Firm
- ESAOTE S.P.A. Sesto Fiorentino Via Degli Olmi 11 Sesto Fiorentino Italy
- FEI Number
- 3012981490
- Product Code
- IYN
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- January 3, 2018
- Terminated
- May 30, 2019
Description
MyLab Ultrasound System; Models 6100, 6150, 6200, 6250; Probes: BL433, C 18, CA541, EC1123, EC123, IH 618, IOT332, IOT342, L 415, LA332E, LA435, LA523, LA533, SI2C41, SL2325, TRT33
The ultrasound system used in certain settings may result in overheating of the ultrasound probe head.
The January 3, 2018, notification letter requests the consignee complete and return a recall response form. The letter states that Esaote customer service will contact the consignee about the software update. In the meantime, it is recommended that the consignee does not use the PA023E probe in Pulsed Wave (PW) Doppler mode and that they refrain from using the system in Elaxto mode where the patient is unconscious or sedated. The consignee is asked to check the temperature of the probe before use.
US distribution in the states: AZ, FL, OK, PA, and TX.
5