Staple Removal Set/Kit; a sterile Rx convenience kit containing a staple remover and a gauze sponge. Theses kits were marketed under the following labels, 50 kits per case: a) Medikmark inc. Staple Removal Set, reorder no. SR-2005, 3600 Bur Wood Drive, Waukegan, IL 60085; b) Maxxim Medical Staple Removal Kit, reorder no. SD-2005, Distributed by Maxxim Medical, Oldsmar, FL 34677.
Recall
- Recall Number
- Z-1138-04
- Event Number
- 29579
- Firm
- Medikmark, Incorporated
- FEI Number
- 1450943
- Product Code
- KKX
- Status
- Terminated
- Root Cause
- Other
- Initiated
- July 12, 2004
- Posted
- July 22, 2004
- Terminated
- June 27, 2006
- Address
- 3600 Burwood Dr, Waukegan, IL, 60085-8399
Description
Staple Removal Set/Kit; a sterile Rx convenience kit containing a staple remover and a gauze sponge. Theses kits were marketed under the following labels, 50 kits per case: a) Medikmark inc. Staple Removal Set, reorder no. SR-2005, 3600 Bur Wood Drive, Waukegan, IL 60085; b) Maxxim Medical Staple Removal Kit, reorder no. SD-2005, Distributed by Maxxim Medical, Oldsmar, FL 34677.
The tip of the staple remover may puncture a hole in the wall of the tray, compromising the sterility of the kit.
Recall notifications were sent by fax and follow-up registered mail to the direct account distributors and users on 7/12-15/04. The accounts were informed of the potential non-sterility of the product and were instructed to immediately discontinue sale/use of the products and place all inventory in quarantine pending return to Medikmark. Distributors were instructed to sub-recall the kits from their customers. Any questions were directed to 1-847-596-7723.
Nationwide
724,800 kits