FDA Recall
Terminated
Immulite 2000. Immunoassay system. Software version 2.6 was to be uploaded on systems running 2.5b software.
Recall: Z-1138-03
·
Initiated June 13, 2003
Recall
- Recall Number
- Z-1138-03
- Event Number
- 26892
- Firm
- DPC Cirrus
- FEI Number
- 1000122204
- Product Code
- JJE
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 13, 2003
- Posted
- August 20, 2003
- Terminated
- November 1, 2004
- Address
- 62 Flanders-Bartley Road, Flanders, NJ, 07836
Description
Immulite 2000. Immunoassay system. Software version 2.6 was to be uploaded on systems running 2.5b software.
Reason
Version 2.6 software was released however it will not properly handle adjustment slopes.
Action
Recall Technical Bulletins were sent out on 6/13/2003 to all Immulite 2000 users instructing them to not install the provided version 2.6 software.
Distribution
Software was distributed to Immulite 2000 users throughout the nation. These units are found in hospitals and clinics. Software was also distributed to the firm''s service representatives.
Quantity
434 upgrades