FDA Recall Terminated

Immulite 2000. Immunoassay system. Software version 2.6 was to be uploaded on systems running 2.5b software.

Recall: Z-1138-03 · Initiated June 13, 2003

Recall

Recall Number
Z-1138-03
Event Number
26892
Firm
DPC Cirrus
FEI Number
1000122204
Product Code
JJE
Status
Terminated
Root Cause
Other
Initiated
June 13, 2003
Posted
August 20, 2003
Terminated
November 1, 2004
Address
62 Flanders-Bartley Road, Flanders, NJ, 07836

Description

Immulite 2000. Immunoassay system. Software version 2.6 was to be uploaded on systems running 2.5b software.

Reason

Version 2.6 software was released however it will not properly handle adjustment slopes.

Action

Recall Technical Bulletins were sent out on 6/13/2003 to all Immulite 2000 users instructing them to not install the provided version 2.6 software.

Distribution

Software was distributed to Immulite 2000 users throughout the nation. These units are found in hospitals and clinics. Software was also distributed to the firm''s service representatives.

Quantity

434 upgrades