Cardinal Health Presource Kits: 1) ACDF PACK, Catalog Number: SNEHBACSCA; 2) ACDF PACK, Catalog Number: SNEHBACSCB
Recall
- Recall Number
- Z-1137-2025
- Event Number
- 95947
- Firm
- Cardinal Health 200, LLC
- FEI Number
- 3001236905
- Product Code
- PVP
- Status
- Open, Classified
- Root Cause
- Material/Component Contamination
- Initiated
- November 8, 2024
- Posted
- February 10, 2025
- Address
- 3651 Birchwood Dr, Waukegan, IL, 60085-8337
Description
Cardinal Health Presource Kits: 1) ACDF PACK, Catalog Number: SNEHBACSCA; 2) ACDF PACK, Catalog Number: SNEHBACSCB
Potential for endotoxin contamination in non-sterile surgical strips and patties in Presource Kits.
An URGENT: MEDICAL DEVICE CORRECTION OVER LABEL notification letter dated 11/8/24 was sent to customers. If you locate affected product within your facility, please complete the following: 1. QUARANTINE affected kits & ALERT your clinicians of this action 2. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard all CODMAN Patties & Strips included with the Presource product. 3. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Should you have any questions, require additional labels or desire special assistance relating to this recall, please feel free to contact Cardinal Health Quality Systems at 800.292.9332.
Worldwide distribution - US Nationwide and the countries of Canada, Saudi Arabia.
973,785 total units