FDA Recall Open, Classified

PHILIPS BATTERY, V60 14.4V, 11.0 Ah, 163 Wh, REF 1076374(1058272), LOT M94663-P1 - Product Usage: The ventilator is intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories.

Recall: Z-1137-2021 · Initiated January 6, 2021

Recall

Recall Number
Z-1137-2021
Event Number
87314
Firm
Respironics California, LLC
FEI Number
2518422
Product Code
MNT
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
January 6, 2021
Address
2271 Cosmos Ct, Carlsbad, CA, 92011-1517

Description

PHILIPS BATTERY, V60 14.4V, 11.0 Ah, 163 Wh, REF 1076374(1058272), LOT M94663-P1 - Product Usage: The ventilator is intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories.

Reason

Due to unqualified "design verification" batteries being shipped to customers.

Action

On 01/06/2021, the firm contacted via telephone calls and then on 01/29/2021 sent a follow-up "URGENT - Medical Device Removal" letter via priority mail to customers to inform them that the Recalling Firm shipped unqualified batteries to customers. These batteries are similar to production batteries but did complete final qualification testing. Potential hazards that may result, In a worst-case scenario, a battery failure could occur while in use on a patient during intra-hospital transport or when not connected to AC power. In both situations, the patient could lose ventilatory assistance and experience severe hypoxemia. Customers are provided instructions on how to identify affected products: Check the label of backup batteries (PN 1076374) received on or after November 25, 2020, or installed on affected units. If the battery has LOT M94663-P1, it is affected. Customers are instructed to: 1. If the batteries have been installed in ventilators, remove the ventilators from service until a field service representative or approved service provider schedules a visit and replaces the battery. 2. If waiting for a field service representative or approved service provider is not possible given their current census, contact Recalling Firm by email at [email protected] to request a replacement battery that trained biomed can install. 3. If the affected battery has not been installed in a ventilator, quarantine the battery until a Recalling Firm representative can replace it. 4. Complete, sign, and return the Acknowledgment and Receipt Form at the end of this letter. For further information and questions - contact the local representatives or Customer Care Solutions Center (24/7) +1-800-722-9377

Distribution

U.S. Nationwide distribution including in the states of FL, GA, IA, IA, IL. IL, MN, MO, NC, ND, OH, TX, TX and UT.

Quantity

94 batteries