SpyScope DS Access and Delivery diagnostic biliary catheter, M00546600
Recall
- Recall Number
- Z-1133-2018
- Event Number
- 79529
- Firm
- Boston Scientific Corporation
- FEI Number
- 3005099803
- Product Code
- FBN
- Status
- Terminated
- Root Cause
- Process design
- Initiated
- December 20, 2017
- Terminated
- December 13, 2019
- Address
- 100 Boston Scientific Way, Marlborough, MA, 01752-1234
Description
SpyScope DS Access and Delivery diagnostic biliary catheter, M00546600
Certain lots of devices may have the working channel sleeve protruding from the camera cap which may result in injuries from minor tissue damage to significant bleeding.
The notification letter requests the consignee to identify and return the affected product along with the response form. Distributors are asked to share the recall information to the customer level. 12/20/2017 by mail.
Worldwide Distribution US nationwide, Austria, Belarus, Belguim, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Hungary, Ireland, Israel, Italy, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Spain, Sweden, Switzerland, Bahrain, China, Hong Kong, India, Malaysia, Pakistan, Qatar, Saudi Arabia, Singapore, South Africa, South Korea, Thailand, Tunisia, Turkey, Vietnam, Brazil, Canada, Cost Rica, Ecuador, Guatemala, Mexico, Peru