TRIGEN (TM) L-P SCREW, 4.5 MM X 37.5 MM, TI-6AL-4V, QTY: (1), STERILE R, REF 71645437, Smith & Nephew, Inc. Memphis, TN 38116 USA orthopedic surgery
Recall
- Recall Number
- Z-1132-2013
- Event Number
- 64669
- Firm
- Smith & Nephew Inc
- FEI Number
- 1020279
- Product Code
- HSB
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- March 6, 2013
- Posted
- April 17, 2013
- Terminated
- August 16, 2019
- Address
- 1450 E. Brooks Rd, Memphis, TN, 38116
Description
TRIGEN (TM) L-P SCREW, 4.5 MM X 37.5 MM, TI-6AL-4V, QTY: (1), STERILE R, REF 71645437, Smith & Nephew, Inc. Memphis, TN 38116 USA orthopedic surgery
One batch of 4.5 mm diameter Trigen Low-Profile Bone Screws were incorrectly anodized with a gold coating (indicating a 5 mm diameter) instead of a gray coating (indicating a 4.5 mm diameter).
Smith & Nephew sent an Urgent Product Recall Notification letter dated March 6, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to cease use and distribution of the affected product and quarantine for return. Non-responding consignees were notified again on March 15, 2013, by email. Surgeons were to be notified by letter beginning on March 20, 2013. For questions regarding this recall call 978-749-1330.
Nationwide Distribution including CA, FL, GA, MA, MI, NC, SC, OH, and MO.
23 units