Clinitron Rite Hite Bed headboard and foot side rails. The Clinitron Rite Hite Air Fluidized Therapy Unit is a patient management system that combines air fluidized therapy and low airloss therapy technologies together on a low-height frame. The lower body section provides air fluidized therapy, and the upper body section provides low airloss therapy.

Recall: Z-1132-2012 · Initiated February 3, 2012

Recall

Recall Number
Z-1132-2012
Event Number
60888
Firm
Hill-Rom, Inc.
FEI Number
1824206
Product Code
INX
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
February 3, 2012
Posted
March 1, 2012
Terminated
December 5, 2012
Address
1069 State Route 46 East, Batesville, IN, 47006-7520

Description

Clinitron Rite Hite Bed headboard and foot side rails. The Clinitron Rite Hite Air Fluidized Therapy Unit is a patient management system that combines air fluidized therapy and low airloss therapy technologies together on a low-height frame. The lower body section provides air fluidized therapy, and the upper body section provides low airloss therapy.

Reason

The firm identified in October 2011 that the Clinitron Rite Hite Bed and Clinitron At Home Bed headboards failed to meet the safety standard force requirements. The firm identified in October 2011 that the Clinitron Rite Hite Bed and Clinitron At Home Bed outer rail, which is extruded Aluminum tubing, is intended to be one piece joined together by a weld at the lower center. The parts receiv

Action

Hill-Rom, Inc. sent a Modification Notice Urgent Medical Device Correction Notification dated February 2, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The modification notice instructed all customers to find and replace all headboards and food side rails that do not meet design intent. All technicians will inspect all affected product in the service centers and all at home and rite hite beds for an "x" or "L" stamp. The technicians will inspect the beds that are still in use during scheduled maintenance or before each rental. All will be completed before 01-JUL-12. For support call 1-800-445-3720

Distribution

Worldwide Distribution -- USA (nationwide) including the states of AL, AK, AZ, AR, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, WA, WI, WV, and WY. and the country of Canada.

Quantity

Foot Siderail = 70; Headboards = 103