FDA Recall Open, Classified

Cardinal Health Presource Kits: 1) LARYNGOSCOPY PACK, Catalog Number:SRIET0979N;

Recall: Z-1131-2025 · Initiated November 8, 2024

Recall

Recall Number
Z-1131-2025
Event Number
95947
Firm
Cardinal Health 200, LLC
FEI Number
3001236905
Product Code
OGH
Status
Open, Classified
Root Cause
Material/Component Contamination
Initiated
November 8, 2024
Posted
February 10, 2025
Address
3651 Birchwood Dr, Waukegan, IL, 60085-8337

Description

Cardinal Health Presource Kits: 1) LARYNGOSCOPY PACK, Catalog Number:SRIET0979N;

Reason

Potential for endotoxin contamination in non-sterile surgical strips and patties in Presource Kits.

Action

An URGENT: MEDICAL DEVICE CORRECTION OVER LABEL notification letter dated 11/8/24 was sent to customers. If you locate affected product within your facility, please complete the following: 1. QUARANTINE affected kits & ALERT your clinicians of this action 2. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard all CODMAN Patties & Strips included with the Presource product. 3. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Should you have any questions, require additional labels or desire special assistance relating to this recall, please feel free to contact Cardinal Health Quality Systems at 800.292.9332.

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Saudi Arabia.

Quantity

973,785 total units