FDA Recall Terminated

ACUVUE OASYS for ASTIGMATISM brand contact lenses, Base Curve 8.6 Refractive Power -1.00D, -1.25D x 100, 6 pack.

Recall: Z-1131-2019 · Initiated March 11, 2019

Recall

Recall Number
Z-1131-2019
Event Number
82402
Firm
Johnson & Johnson Vision Care, Inc.
FEI Number
1000222023
Product Code
LPL
Status
Terminated
Root Cause
Equipment maintenance
Initiated
March 11, 2019
Posted
April 11, 2019
Terminated
December 6, 2021
Address
7500 Centurion Pkwy, Jacksonville, FL, 32256-0517

Description

ACUVUE OASYS for ASTIGMATISM brand contact lenses, Base Curve 8.6 Refractive Power -1.00D, -1.25D x 100, 6 pack.

Reason

One lot of ACUVUE OASYS for ASTIGMATISM Brand Contact lenses has the potential for a limited number of individual contact lens packages to have an incomplete packaging seal.

Action

Johnson & Johnson Vision Care, Inc. notified the single distributor on about 03/11/2019 via "URGENT ACTION: VOLUNTARY and LIMITED MEDICAL DEVICE RECALL" letter. The letter instructed the consignee to review their inventory to determine if any affected contact lenses are on hand, stop using all affected lenses, provide the recall notification to anyone within the organization that should be aware of the issue, and notify any customers may have received any affected product and request they discontinue use and return any on hand. They advised the consignee to contact Customer Service at 1-800-843-2020 to arrange for the return and replacement of affected lenses, and they were requested to complete and return the Customer Reply Form.

Distribution

Distributed to California.

Quantity

300 lenses