Haemonetics TEG Hemostasis System Level II Control, Catalog number: 8002. Level II Control is a biological control for use with the Thrombelastograph (TEG) Hemostasis System.

Recall: Z-1130-2015 · Initiated January 26, 2015

Recall

Recall Number
Z-1130-2015
Event Number
70352
Firm
Haemonetics Corporation
FEI Number
1219343
Product Code
JPA
Status
Terminated
Root Cause
Process control
Initiated
January 26, 2015
Posted
February 19, 2015
Terminated
June 5, 2015
Address
400 Wood Rd, Braintree, MA, 02184-2412

Description

Haemonetics TEG Hemostasis System Level II Control, Catalog number: 8002. Level II Control is a biological control for use with the Thrombelastograph (TEG) Hemostasis System.

Reason

Specific lots of Level II Control Kits produce high Maximum Amplitude (MA) results, which are outside of the manufacturer's expected Quality Control range for MA parameter.

Action

Haemonetics issued on 1/26/15, an Urgent Medical Device Recall to consignees The letter identified the affected product, the reason for the recall, as well as the risk to health. It instructs consignees to inspect their stock and determine if they have any of the affected lots. Customers are to immediately discontinue use and discard as instructed any affected product. The attached customer acknowledgment form should be completed and returned whether or not customers have any affected product. Customers are to contact their local service representative at 1-800-537-2802 if replacement product is needed. If any affected product has been forwarded to other locations, customers are to inform them of the recall. Please call the Technical Support team at 1-800-438-2834, with questions about this action.

Distribution

Worldwide Distribution -- US, Canada, Czech Republic, Denmark, Great Britain, and Sweden.

Quantity

4,143 kits