Cleaning probe included as an accessory to aspirators that are sterile packed single use disposable accessories to the GE InstaTrak 3500 Plus x-ray system. Probes are packaged and shipped together with the following aspirators: Straight Aspirator - 1000370-NAV, 7 French Aspirator - 100936, Keane - Moses Extended Straight Aspirator - 1000748, 15 degree aspirator - 1003895, Fried 45 Degree Aspirator - 1000541, Kuhn 90 degree Aspirator - 1000543-NAV.
Recall
- Recall Number
- Z-1130-2007
- Event Number
- 38373
- Firm
- GE OEC Medical Systems, Inc
- FEI Number
- 1720753
- Product Code
- LLZ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- May 18, 2007
- Posted
- August 7, 2007
- Terminated
- August 22, 2011
- Address
- 384 Wright Brothers Dr, Salt Lake City, UT, 84116-2862
Description
Cleaning probe included as an accessory to aspirators that are sterile packed single use disposable accessories to the GE InstaTrak 3500 Plus x-ray system. Probes are packaged and shipped together with the following aspirators: Straight Aspirator - 1000370-NAV, 7 French Aspirator - 100936, Keane - Moses Extended Straight Aspirator - 1000748, 15 degree aspirator - 1003895, Fried 45 Degree Aspirator - 1000541, Kuhn 90 degree Aspirator - 1000543-NAV.
Cleaning probes used to unclog disposable aspirators may shed small particles of nylon material which may be transferred into the nasal cavity.
Consignees were notified by certified mail on 05/18/2007 and were told to stop using and to dispose of the affected probes.
Worldwide, including USA, Canada and France.
67,625 units