FDA Recall Terminated

LifeScan brand OneTouch Ultra Control Solution, Part Numbers/Model Numbers: 01045808, 02141603. Testing of Blood Glucose Levels.

Recall: Z-1129-2013 · Initiated February 25, 2013

Recall

Recall Number
Z-1129-2013
Event Number
64478
Firm
Lifescan Inc
FEI Number
2939301
Product Code
JJX
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
February 25, 2013
Posted
April 17, 2013
Terminated
October 23, 2013
Address
1000 Gibraltar Dr, Milpitas, CA, 95035-6301

Description

LifeScan brand OneTouch Ultra Control Solution, Part Numbers/Model Numbers: 01045808, 02141603. Testing of Blood Glucose Levels.

Reason

When OneTouch Ultra and OneTouch Select test strips are tested with Ultra and Select control solution the meter may display a result outside the control solution test range printed on the strip vial label even though the system is working properly.

Action

LifeScan sent an "URGENT MEDICAL DEVICE RECALL" letter dated February 25, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Customers may contact OneTouch Healthcare Professional Line at 866-461-0089 for questions about this notice.

Distribution

Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, 10, IL, IN, KS, KY, LA, MA, MD, ME, Ml, MN, MO, MS, MT, NC, NO, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SO, TN, TX, UT, VA, VI, VT, WA, WI,WV, and WY, and the countries of Russia and Ukraine.

Quantity

132,301 units total