FDA Recall Terminated

Restoris Partial Knee Application (PKA) RIO (TGS 2.0). For use with the Robotic Arm Interactive Orthopedic System (RIO), is indicated for use in surgical knee procedures.

Recall: Z-1128-2015 · Initiated January 22, 2015

Recall

Recall Number
Z-1128-2015
Event Number
70374
Firm
Mako Surgical Corporation
FEI Number
3005985723
Product Code
OLO
Status
Terminated
Root Cause
Device Design
Initiated
January 22, 2015
Posted
February 19, 2015
Terminated
December 14, 2016
Address
2555 Davie Rd, Ste 110, Plantation, FL, 33317-7424

Description

Restoris Partial Knee Application (PKA) RIO (TGS 2.0). For use with the Robotic Arm Interactive Orthopedic System (RIO), is indicated for use in surgical knee procedures.

Reason

When using the MAKOplasty partial knee Arthroplasty application, the burr continues spinning outside of the stereotactic boundary and after the control switches (foot pedal and trigger) cease to be activated

Action

On January 22, 2015 the firm sent an IMPORTANT MEDICAL DEVICE CORRECTION asking consignees to complete the attached acknowledgement form and return them by email, fax or mail. The notice also noted the RIO system can continue to be used. The likelihood of occurrence is rare (.035% of cases) and the event is readily detectable and can be resolved. Should you have any questions regarding this Important Medical Device Recall Notice, to contact us at (954) 628-1721.

Distribution

Nationwide Distribution-including the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, LA, MA, MD, MI, MO, MS, NC, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV and Hawaii.

Quantity

228 units