FDA Recall Terminated

SR Instruments FG3105 Patient Scales Used on various Joerns Healthcare Patient Lifters. Model# FG3105 Patient Scales. Product Usage: The FG3105 Scale is designed for Joerns Healthcare for use in portable patient weighing systems. The scale is designed to be connected between the patient hoist mechanism and patient sling apparatus. The scale is not intended to be used for patient transport.

Recall: Z-1128-2013 · Initiated December 13, 2012

Recall

Recall Number
Z-1128-2013
Event Number
64698
Firm
Sr Instruments Inc
FEI Number
1317256
Product Code
FRW
Status
Terminated
Root Cause
Use error
Initiated
December 13, 2012
Posted
April 17, 2013
Terminated
March 12, 2015
Address
600 Young St, Tonawanda, NY, 14150

Description

SR Instruments FG3105 Patient Scales Used on various Joerns Healthcare Patient Lifters. Model# FG3105 Patient Scales. Product Usage: The FG3105 Scale is designed for Joerns Healthcare for use in portable patient weighing systems. The scale is designed to be connected between the patient hoist mechanism and patient sling apparatus. The scale is not intended to be used for patient transport.

Reason

SR Instruments Inc. is implementing a field correction for the Joerns Healthcare Patient Lifts after they received two incident reports regarding two field failures of a tensile type load cell involving the FG3105 Patient Scale. The failures resulted in patient falls.

Action

SR Instruments, Inc. sent a Notification letter dated December 13, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. For questions call 716-693-5977.

Distribution

USA Nationwide Distribution in the state of Texas.

Quantity

806 patient scales