FDA Recall
Terminated
Philips Medical HeartStartXL Defibrillator/Monitor Model : M4735A
Recall: Z-1128-2007
·
Initiated June 11, 2007
Recall
- Recall Number
- Z-1128-2007
- Event Number
- 38239
- Firm
- Philips Medical Systems
- FEI Number
- 1218950
- Product Code
- MKJ
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- June 11, 2007
- Posted
- August 7, 2007
- Terminated
- August 8, 2016
- Address
- 3000 Minuteman Road, Andover, MA, 01810-1032
Description
Philips Medical HeartStartXL Defibrillator/Monitor Model : M4735A
Reason
The ECG signal may become unusable to assess patient condition and need for therapy.
Action
Philips Medical contacted accounts by letter dated June 11, 2007. Users are advised to remove product from service, if possible. If not, a back up unit should be available to use. Service reps will contact accounts to replace units.
Distribution
Worldwide, including USA, Canada, Australia, Brazil, China, Czech Republic, France, Germany, Hong Kong, India, Ireland, Italy, Japan, Korea, Kuwait, Macau, Poland, Portugal, Romania, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, UK, and Venezuela.
Quantity
312 units