Accu-Chek FlexLink 1 infusion set; 8/30; Ref number 04540778001. (Distributed outside of the U.S. only).
Recall
- Recall Number
- Z-1128-06
- Event Number
- 34905
- Firm
- Disetronic Medical Systems, Inc.
- FEI Number
- 1000116163
- Product Code
- LZG
- Status
- Terminated
- Root Cause
- Other
- Initiated
- March 31, 2006
- Posted
- July 8, 2006
- Terminated
- November 6, 2007
- Address
- 11800 Exit 5 Parkway, Suite 120, Fishers, IN, 46038
Description
Accu-Chek FlexLink 1 infusion set; 8/30; Ref number 04540778001. (Distributed outside of the U.S. only).
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
"U.S. and International distributors, patients and their physicians were notified via letters dated 3/31/06. Distributors were instructed to search for all part numbers affected by the recall and to provide all affected customers and sub-distributors with the recall notice. Customers were offered a replacement set or replacement with a different model. A press release was also issued on 4/3/06.
Worldwide including USA, Argentina, Austria, Barbados, Belgium, Canada, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Honduras, Hungary, Italy, Mexico, Netherlands, New Zealand, Norway, Poland, Portugal, Singapore, Slovenia, Spain, Sweden, Switzerland and United Kingdom.
36,500 outside North America