FDA Recall Terminated

Easy Access Software releases 1.1, 5.2, 6.2 & 7.2

Recall: Z-1125-03 · Initiated June 27, 2003

Recall

Recall Number
Z-1125-03
Event Number
26514
Firm
Philips Medical Systems
FEI Number
1000524572
Product Code
LLZ
Status
Terminated
Root Cause
Other
Initiated
June 27, 2003
Posted
August 12, 2003
Terminated
January 12, 2005
Address
22100 Bothell Everett Highway, Bothell, WA, 98041

Description

Easy Access Software releases 1.1, 5.2, 6.2 & 7.2

Reason

If DICOM fields for body part, description or accession number are changed, change won't update on the EasyWeb client or hospital systems.

Action

On 6/25/03 the firm issued a letter dated 6/15/03 to the affected customers. The letter advises of the problem and advises the customer that the firm will install corrected software in affected units.

Distribution

Devices were distributed to 23 hospitals and medical centers throughout the US.

Quantity

46 devices