Accu-Chek Ultraflex-2 infusion set; 8 mm 43'' (110 cm); Catalog number INF 04540948001.
Recall
- Recall Number
- Z-1124-06
- Event Number
- 34905
- Firm
- Disetronic Medical Systems, Inc.
- FEI Number
- 1000116163
- Product Code
- LZG
- Status
- Terminated
- Root Cause
- Other
- Initiated
- March 31, 2006
- Posted
- July 8, 2006
- Terminated
- November 6, 2007
- Address
- 11800 Exit 5 Parkway, Suite 120, Fishers, IN, 46038
Description
Accu-Chek Ultraflex-2 infusion set; 8 mm 43'' (110 cm); Catalog number INF 04540948001.
The luer tube may break at the connection to the pump and, if this is not noticed by the patient, the patient may receive an under delivery of insulin.
"U.S. and International distributors, patients and their physicians were notified via letters dated 3/31/06. Distributors were instructed to search for all part numbers affected by the recall and to provide all affected customers and sub-distributors with the recall notice. Customers were offered a replacement set or replacement with a different model. A press release was also issued on 4/3/06.
Worldwide including USA, Argentina, Austria, Barbados, Belgium, Canada, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Honduras, Hungary, Italy, Mexico, Netherlands, New Zealand, Norway, Poland, Portugal, Singapore, Slovenia, Spain, Sweden, Switzerland and United Kingdom.
28,170 in US and Canada