AMO TECNIS CL Foldable Silicone IOL with OptiEdge Design, Model Z9002, diopter size: 16.0 and 21.0. Indicated for primary implantation for the visual correction of aphakia in adult patients.
Recall
- Recall Number
- Z-1123-2014
- Event Number
- 67385
- Firm
- AMO Puerto Rico Manufacturing, Inc.
- FEI Number
- 2648035
- Product Code
- HQL
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- January 28, 2014
- Posted
- March 3, 2014
- Terminated
- April 19, 2017
- Address
- Rd # 402, Km 4.2, Anasco, PR, 00610
Description
AMO TECNIS CL Foldable Silicone IOL with OptiEdge Design, Model Z9002, diopter size: 16.0 and 21.0. Indicated for primary implantation for the visual correction of aphakia in adult patients.
Certain Silicone TECNIS IOLs may be labeled with the incorrect diopter power.
Beginning on January 28, 2014, AMO began distribution of the Recall Notification Letter to the 72 AMO TECNIS IOL customer accounts within the United States who had received lenses from the 2 affected P/O's. 100% delivery of the recall communication to US customers was confirmed on January 31, 2014. Included with the Recall Notification Letter is a facsimile delivery confirmation which each customer is instructed to fax back to AMO.
Worldwide Distribution-USA (nationwide) and the countries of Australia, Great Britain, Canada and Japan.
169 units