FDA Recall Terminated

ADVIA Centaur HAV IgM, 100 Test Kit, used on the ADVIA Centaur or XP, REF: 05004126, in vitro diagnostic Hepatitis A test, Siemens Medical Solutions Diagnostics, Tarrytown, NY 10591.

Recall: Z-1119-2008 · Initiated January 12, 2008

Recall

Recall Number
Z-1119-2008
Event Number
46346
Firm
Siemens Medical Solutions Diagnostics
FEI Number
1219913
Product Code
LOL
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
January 12, 2008
Posted
May 6, 2008
Terminated
June 20, 2012
Address
333 Coney St, East Walpole, MA, 02032-1516

Description

ADVIA Centaur HAV IgM, 100 Test Kit, used on the ADVIA Centaur or XP, REF: 05004126, in vitro diagnostic Hepatitis A test, Siemens Medical Solutions Diagnostics, Tarrytown, NY 10591.

Reason

Incorrect results: False reactive specimens on the ADVIA Centaur HAV IgM assay.

Action

Siemens issued a Urgent Field Safety Notiifcation (078D0709-01) to all Siemens Diagnostics Branches on 1/12/08 for communication to their customers. This Customer Bulletin/Urgent Field Safety Notification advises the customer of the problem and the actions to be taken to mitigate risk. A Confirmation fax-back form is included to ensure that all customers worldwide have been notified and have complied with the mandatory action.

Distribution

Worldwide Distribution - USA, Puerto Rico, Argentina, Australia, Brazil, Canada, Chile, Colombia, Czech Republic, Egypt, France, Germany, Greece, Hong Kong, India, Israel, Italy, Malaysia, Mexico, the Netherlands, New Zealand, Norway, Peru, Poland, Portugal, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, Uruguay, and Venezuela.

Quantity

49,533