FDA Recall Terminated

Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery.

Recall: Z-1116-2015 · Initiated February 6, 2015

Recall

Recall Number
Z-1116-2015
Event Number
70492
Firm
Terumo Cardiovascular Systems Corporation
FEI Number
1828100
Product Code
DWC
Status
Terminated
Root Cause
Device Design
Initiated
February 6, 2015
Posted
February 13, 2015
Terminated
October 17, 2016
Address
6200 Jackson Rd, Ann Arbor, MI, 48103-9586

Description

Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery.

Reason

Malfunctioning mixing valve and loss of cooling and/or heating capability. System goes to standby mode and shows error messages EOd (invalid mix valve measured range of motion) or EOE (valve unable to move or complete a motion). Mixing valves need to be replaced at regular intervals.

Action

On 2/6/2015, URGENT MEDICAL DEVICE RECALL notifications were mailed to the affected hospitals. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call customer service at 1-800-521-2818.

Distribution

US (nationwide) including the states of AR, CA, CO, CT, FL, IA, IL, KS, KY, MD, MI, MO, MS, NY, OH, PA, TX, UT, WA, and WI and the country of Japan.

Quantity

103 units