FDA Recall Terminated

Perkin Elmer Genetic Screening Processor (GSP) Product code: 2021-0010. It is intended for in vitro quantitative / qualitative determination of analytes in body fluids.

Recall: Z-1116-2012 · Initiated January 9, 2012

Recall

Recall Number
Z-1116-2012
Event Number
61031
Firm
PerkinElmer Life and Analytical Sciences, Wallac, OY Mustionkatu 6
FEI Number
3002808498
Product Code
KHO
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
January 9, 2012
Posted
February 28, 2012
Terminated
March 6, 2013
Address
P.O. Box 10, Turku Finland

Description

Perkin Elmer Genetic Screening Processor (GSP) Product code: 2021-0010. It is intended for in vitro quantitative / qualitative determination of analytes in body fluids.

Reason

GSP instruments with defective washer modules may cause a false negative screening result being reported for T4, TSH and IRT. For 17-OHP, a false positive result may be generated if the failure occurs for a normal sample.

Action

Perkin Elmer issued an "Urgent Medical Device Recall" letter dated 1/25/12 to all customers and informed them about the possible issue and requested permission for service personnel to check their instrument. The customers were instructed that the instrument could not be used and the washer module required replacement. Customers were told to expect to be contacted by the firm's local representative for further details.

Distribution

Worldwide Distribution - USA, including the states of Florida, Ohio, Oregon, Texas and the countries of Brazil, Belgium, France, Germany, and the UK.

Quantity

10 units