FDA Recall
Terminated
Hitachi AIRIS II MRI System
Recall: Z-1115-06
·
Initiated April 25, 2006
Recall
- Recall Number
- Z-1115-06
- Event Number
- 35107
- Firm
- Hitachi Medical Systems America Inc
- FEI Number
- 1528028
- Product Code
- LNH
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 25, 2006
- Posted
- June 14, 2006
- Terminated
- October 29, 2008
- Address
- 1959 Summit Commerce Park, Twinsburg, OH, 44087-2371
Description
Hitachi AIRIS II MRI System
Reason
Improper Assembly-The AIRIS/AIRIS II MRI system has the potential for the input wiring of a power transformer to overheat due to damaged wiring or possible loose connections.
Action
The recalling firm plans to inspect the transformer connections for damaged wiring and check the connections for tightness and tighten as necessary. The inspections and possible corrections began on 4/25/2006 and are in the process of being conducted by the firm''s Field Service Engineers. The firm anticipates trhat these actions will be completed by 6/25/2006.
Distribution
The devices were distributed to consignees located throughout the United States.
Quantity
248