FDA Recall Terminated

Hitachi AIRIS II MRI System

Recall: Z-1115-06 · Initiated April 25, 2006

Recall

Recall Number
Z-1115-06
Event Number
35107
Firm
Hitachi Medical Systems America Inc
FEI Number
1528028
Product Code
LNH
Status
Terminated
Root Cause
Other
Initiated
April 25, 2006
Posted
June 14, 2006
Terminated
October 29, 2008
Address
1959 Summit Commerce Park, Twinsburg, OH, 44087-2371

Description

Hitachi AIRIS II MRI System

Reason

Improper Assembly-The AIRIS/AIRIS II MRI system has the potential for the input wiring of a power transformer to overheat due to damaged wiring or possible loose connections.

Action

The recalling firm plans to inspect the transformer connections for damaged wiring and check the connections for tightness and tighten as necessary. The inspections and possible corrections began on 4/25/2006 and are in the process of being conducted by the firm''s Field Service Engineers. The firm anticipates trhat these actions will be completed by 6/25/2006.

Distribution

The devices were distributed to consignees located throughout the United States.

Quantity

248