Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems
Recall
- Recall Number
- Z-1112-2025
- Event Number
- 96089
- Firm
- Beckman Coulter Inc.
- FEI Number
- 2050012
- Product Code
- PTF
- Status
- Open, Classified
- Root Cause
- Under Investigation by firm
- Initiated
- January 2, 2025
- Posted
- February 5, 2025
- Address
- 250 S Kraemer Blvd, Brea, CA, 92821-6232
Description
Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems
Identified reagent lots are experiencing a high rate of calibration failures with failure code of "Max Iterate" on Access 2 and UniCel Dxl Immunoassay Systems. Reporting of results may be delayed due to the time required to achieve a passing calibration curve.
An Urgent Medical Device Recall letter was sent during the week of December 30, 2024, to the affected customers via first class mail and email. If a passing calibration curve can be generated using the Access PCT reagent lots listed above, no action is needed. If repeated calibration failures with a failure code of Max Iterate are obtained using the Access PCT reagent lots listed above, discontinue using the affected reagent lot number.
Worldwide distribution - US Nationwide and the countries of Austria, Bahrain, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Croatia, Czechia, Denmark, France, Germany, Greece, Hungary, India, Ireland, Italy, Kazakhstan, Korea, Republic of, Malaysia, Mexico, Peru, Philippines, Poland, Puerto Rico, Romania, Slovakia, South Africa, Spain, Switzerland, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Viet Nam.
6,994 US; 10,067 OUS