Tego Connector: REF: 011-D1000, 01C-D1000, 055-D1000, D1000; Conector Tego: REF: LAT-D1000
Recall
- Recall Number
- Z-1111-2026
- Event Number
- 98150
- Firm
- ICU Medical, Inc.
- FEI Number
- 2025816
- Product Code
- FPA
- Status
- Open, Classified
- Root Cause
- Process control
- Initiated
- December 19, 2025
- Posted
- January 16, 2026
- Address
- 951 Calle, Amanecer San Clemente, CA, 92673-6212
Description
Tego Connector: REF: 011-D1000, 01C-D1000, 055-D1000, D1000; Conector Tego: REF: LAT-D1000
Needle free access device intended as accessory to vascular access used in hemodialysis or accessory to intravascular administration sets has a silicone seal which may bulge at the top, separate from the body, tear; may result in fluid leaks, occluded fluid flow, inability to inject/withdraw blood product, delayed/interrupted therapy, exposure to biological contaminants, air infused into the body
On 12/19/2025, correction notices were emailed to customers who were asked to do the following: 1) Destroy all affected products following your institution's process for destruction. If destroying is not immediately possible at your facility, then the product should be quarantined until disposal. 2) Share this notification with all potential users of the device, to ensure they are aware of this notification. If the devices are used at another location, please ensure this communication is delivered there. 3) Complete and return the Customer Response Form via email to [email protected] For further inquiries, please contact the firm: Global Complaint Management: [email protected] 1-(866)-216-8806; Customer Service: [email protected], 1-(800)-258-5361
Worldwide - US Nationwide distribution including in the states of FL, AZ, RI, TN, KS, NJ, NY, VA, SC, GA, CA, MS, TX, OH, NC, MA, WA, IL, PA, NV, NH, LA, MI, NE, ME, MN, NM, OK, MD, IA, OR, SD, WI, MO, AL, WY, IN, KY, DC, UT, AR, WV, AK, ND, CO and the countries of BE, DE, CO, ES, AU, AR, SA, SI, IT, LK, MY, AT, FI, GB, TW, MX, NL, IE, CA, GR, ZA, SE, SK, NZ, LT, CH, HR, FR, JO, PE, SG, BR, PL, RO, HU, CZ, NO, AE, EE, CN, MA, HK.
14,684,607 items