FDA Recall Terminated

Cascade Abrazo aPTT Test Card - Product Usage: are to be used with the Cascade Abrazo analyzer and are intended for the determination of the activated Partial Thromboplastin Time (aPTT) of citrated whole blood or plasma, Model 5722, For Export Only.

Recall: Z-1108-2021 · Initiated January 5, 2021

Recall

Recall Number
Z-1108-2021
Event Number
87151
Firm
Helena Laboratories, Corp.
FEI Number
1618982
Product Code
GFO
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
January 5, 2021
Terminated
November 19, 2024
Address
1530 Lindbergh Dr, Beaumont, TX, 77707-4131

Description

Cascade Abrazo aPTT Test Card - Product Usage: are to be used with the Cascade Abrazo analyzer and are intended for the determination of the activated Partial Thromboplastin Time (aPTT) of citrated whole blood or plasma, Model 5722, For Export Only.

Reason

Intermittent potential for shorter than expected clot times.

Action

On 01/05/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification Letter via email to it distributors informing them that this Recall has been identified due to the potential for shorter than expected clot times on abnormal patients and on abnormal controls on some cards from specific lots and the use of these products may include difficulty achieving abnormal control results in range and incorrectly low patient results. Distributors were instructed to: 1) Immediately examine their inventory and quarantine the specific product lots 2) If they have further distributed the affected product, they are to identify these customers and notify them at once. In addition, the notification to their customers may be enhanced by including a copy of the recall notification letter, which must include instruction of what customers should do with the recalled product. 3) Complete and return enclosed response form as soon as possible Any questions, email or contact Dr. Jessica Jones Hanka at [email protected] or 1-800-231-5663 ext. 1177. Distributors were contacted by email and requested to evaluate inventory while the Recalling Firm investigated issue . Distributors were formally notified of recall on 1/5/2021 and provided with a template for a notification letter to translate and send to end users in each country. End users will be asked to return unused product to the distributor, and distributors will forward all returned product to Helena Labs. Distributors will be asked to provide an account of all product distributed. Effectiveness will be determined by evaluating the total product returned and the total product accounted for based on information provided by distributors.

Distribution

Worldwide distribution - U.S. Nationwide distribution and the countries of Germany, Italy, Romania, Australia, Chile.

Quantity

205 total kits