FDA Recall Terminated

SIEMENS AXIOM Artis - AUTOMAP FUNCTION

Recall: Z-1108-04 · Initiated June 2, 2004

Recall

Recall Number
Z-1108-04
Event Number
29327
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
IZI
Status
Terminated
Root Cause
Other
Initiated
June 2, 2004
Posted
July 20, 2004
Terminated
October 21, 2004
Address
51 Valley Stream Pkwy, Malvern, PA, 19355-1406

Description

SIEMENS AXIOM Artis - AUTOMAP FUNCTION

Reason

Error appears in the AUTOMAP function, Module 2, when PERISTEEPING or PERIVISION is activated simultaneously that causes system to crash.

Action

The recalling firm sent recall letter to their customers informing them of the software problem. The customers were informed that a Siemens Service Representative will visit each location to deactivate Module 2 of the AUTOMAP function per Update AX023/04S.

Distribution

The recalled product has been distributed throughout the United States, such as: AK, AL, AZ, CA, DC, FL, GA, IA, IL, IN, KY, LA, MI, MN, MO, MS, NC, NJ, NY, OH, OK, PA, SC, TN, TX, VA, WA, WI,

Quantity

127 units