FDA Recall
Terminated
SIEMENS AXIOM Artis - AUTOMAP FUNCTION
Recall: Z-1108-04
·
Initiated June 2, 2004
Recall
- Recall Number
- Z-1108-04
- Event Number
- 29327
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- IZI
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 2, 2004
- Posted
- July 20, 2004
- Terminated
- October 21, 2004
- Address
- 51 Valley Stream Pkwy, Malvern, PA, 19355-1406
Description
SIEMENS AXIOM Artis - AUTOMAP FUNCTION
Reason
Error appears in the AUTOMAP function, Module 2, when PERISTEEPING or PERIVISION is activated simultaneously that causes system to crash.
Action
The recalling firm sent recall letter to their customers informing them of the software problem. The customers were informed that a Siemens Service Representative will visit each location to deactivate Module 2 of the AUTOMAP function per Update AX023/04S.
Distribution
The recalled product has been distributed throughout the United States, such as: AK, AL, AZ, CA, DC, FL, GA, IA, IL, IN, KY, LA, MI, MN, MO, MS, NC, NJ, NY, OH, OK, PA, SC, TN, TX, VA, WA, WI,
Quantity
127 units