FDA Recall Terminated

LIFEPAK 12 defibrillator/monitor Part Numbers: VLP12-02-000000 through VLP12-02-005003, AVLP12-02-000000 through AVLP12-02-005003, DVLP12-02-000000 through DVLP12-02-005003, EVLP12-02-000000 through EVLP12-02-005003, FVLP12-02-000000 through FVLP12-02-005003, LVLP12-02-000000 through LVLP12-02-005003, PVLP12-02-000000 through PVLP12-02-005003, UVLP12-02-000000 through UVLP12-02-005003

Recall: Z-1107-04 · Initiated May 5, 2004

Recall

Recall Number
Z-1107-04
Event Number
29025
Firm
Medtronic Physio Control Corp
FEI Number
3015876
Product Code
MKJ
Status
Terminated
Root Cause
Other
Initiated
May 5, 2004
Posted
July 20, 2004
Terminated
May 6, 2006
Address
11811 Willows Rd NE, Redmond, WA, 98073

Description

LIFEPAK 12 defibrillator/monitor Part Numbers: VLP12-02-000000 through VLP12-02-005003, AVLP12-02-000000 through AVLP12-02-005003, DVLP12-02-000000 through DVLP12-02-005003, EVLP12-02-000000 through EVLP12-02-005003, FVLP12-02-000000 through FVLP12-02-005003, LVLP12-02-000000 through LVLP12-02-005003, PVLP12-02-000000 through PVLP12-02-005003, UVLP12-02-000000 through UVLP12-02-005003

Reason

Therapy cable doesn't meet firm's standards/may not maintain connections sufficient to support ECG monitoring, external pacing or defibrillation; ECG connector with higher than normal connection resistance resulting in intermittent ECG signal; therapy connector has potential for impact damage.

Action

The firm reviewed records and/or contacted customers by telephone to verify affected devices used in the out of hospital mobile environment. All affected customers were notified by a letter dated May 2004. The firm''s service representative will contact customers to inspect and correct the devices.

Distribution

The firm distributes to distributors, hospitals, EMT-Ds and paramedics nationwide. The firm also distibutes the product internationally.

Quantity

20,676 domestic, 19,899 international