FDA Recall Terminated

Bemer-Set/ Pro-Set for the following models: REF. 410201 (BEMER Pro-Set), REF. 420201 (B.BOX Professional), REF. 420100 (B.BOX CLASSIC), REF. 410100 (BEMER Classic-Set), REF. 420200 (B.BOX Professional), and REF. 410200 (BEMER Pro - Set)

Recall: Z-1105-2021 · Initiated January 13, 2021

Recall

Recall Number
Z-1105-2021
Event Number
87094
Firm
BEMER INT. AG Austrasse 15 Triesen Liechtenstein
FEI Number
3009102830
Product Code
NGX
Status
Terminated
Root Cause
Labeling Change Control
Initiated
January 13, 2021
Terminated
July 12, 2023

Description

Bemer-Set/ Pro-Set for the following models: REF. 410201 (BEMER Pro-Set), REF. 420201 (B.BOX Professional), REF. 420100 (B.BOX CLASSIC), REF. 410100 (BEMER Classic-Set), REF. 420200 (B.BOX Professional), and REF. 410200 (BEMER Pro - Set)

Reason

There potential that electromagnetic fields generated in the therapy system may cause technical defects in a patient using an insulin pumps or a malfunction of a pacemaker.

Action

On January 13, 2021, Bemer USA sent a "Urgent Safety Information" letter to affected consignees via E-Mail. In addition to informing consignees about the recall the firm asked consignees to take the following actions: 1. Patients who wear an active medical implant that leads to stimulation (e.g., pacemakers, defibrillators, brain stimulators, muscle stimulators) should not use the BEMER therapy system unless a specialist physician carries out a personal risk assessment and subsequently expressly tells the patient that he/she can continue the treatment. If you are an affected patient, please stop using the BEMER therapy system immediately and consult your specialist physician. The specialist physician will then decide whether or not you can continue using the BEMER therapy system. 2. In the risk assessment, your specialist physician must estimate whether or not the BEMER therapy can cause possible interference at a maximum flow density of 150 T (average flow density for total body treatment 21 T), based on the individually set threshold values of the implant and under consideration of the distance between the applicator (coil) and the pacemaker with its electrodes. If the specialist physician estimates that interference and therefore, operating safety, cannot be excluded, do not continue the BEMER therapy system treatment. 3. Patients wearing active medical implants designed to deliver medication (e.g., medication pumps) have an absolute contraindication. In other words, they may no longer use the BEMER therapy system. These contraindications must be observed by the user for future use. 4. If you have re-sold the aforementioned products in new condition to a third party, you are required to forward a copy of this notice to all your customers, including all professional users (doctors, therapists, hospitals, etc.). Alternatively, please inform the contact person listed below so we can ensure the compliance with this notice requirement. Your applicabl

Distribution

US

Quantity

197,724 units