Cook Central Venous Catheter Tray, Triple Lumen Polyurethane, 5.0Fr/12cm, CUTLMY501JRSCCCT, G26945, Sterile, Cook Medical, Cook Incorporated, Bloomington, IN.
Recall
- Recall Number
- Z-1105-2011
- Event Number
- 57663
- Firm
- Cook, Inc.
- FEI Number
- 1820334
- Product Code
- FOZ
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- October 4, 2010
- Posted
- February 23, 2011
- Terminated
- June 21, 2011
- Address
- 750 Daniels Way, Bloomington, IN, 47404-9120
Description
Cook Central Venous Catheter Tray, Triple Lumen Polyurethane, 5.0Fr/12cm, CUTLMY501JRSCCCT, G26945, Sterile, Cook Medical, Cook Incorporated, Bloomington, IN.
Cook, Inc. is recalling Single, Double, Triple, and Five-Lumen Central Venous Catheter Trays AND Single and Double Lumen PICC Peripherally Inserted Central Venous Catheter Trays. These kits and trays contain 0.9% Sodium Chloride Injection USP 5 mL fill in 6 mL flush syringes that have been recalled by Excelsior Medical. During a routine test for syringe sterility, a possible integrity failure wa
Cook, Inc. sent URGENT DEVICE RECALLS letters to its customers dated 10/4/2010. The letters asked customers to examine their inventory and notify Cook for product return. The letter included a response form.
Worldwide distribution: USA, Chile, Colombia, Panama, Barbados, Canada, Guatemala, and Saudi Arabia.
51,473 total kits/trays