FDA Recall
Terminated
Smith & Nephew TRIVEX System Pneumatic Footswitch, Part #7209791; distributed by Smith & Nephew, Inc. - Endoscopy, Andover, MA 01810.
Recall: Z-1105-2007
·
Initiated October 12, 2004
Recall
- Recall Number
- Z-1105-2007
- Event Number
- 37614
- Firm
- Smith & Nephew, Inc Endoscopy Division
- FEI Number
- 1643264
- Product Code
- DWQ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- October 12, 2004
- Posted
- August 2, 2007
- Terminated
- August 2, 2007
- Address
- 76 S. Meridian Ave, Oklahoma City, OK, 73107-6512
Description
Smith & Nephew TRIVEX System Pneumatic Footswitch, Part #7209791; distributed by Smith & Nephew, Inc. - Endoscopy, Andover, MA 01810.
Reason
Build-up of static charge in Pneumatic Footswitch tubing may cause control unit to reset to default settings in the operating room during surgery.
Action
Firm sent notification letters to consignees via overnight FedEx on 10/12-13/04. International customers were notified via letter on 11/10/04. Customers were asked to return footswitches for a replacement.
Distribution
Product sent to distributors nationwide and to the following countries: Australia, Austria, Canada, China, Germany, Italy, Japan, and Taiwan.
Quantity
274 units.