FDA Recall Terminated

Smith & Nephew TRIVEX System Pneumatic Footswitch, Part #7209791; distributed by Smith & Nephew, Inc. - Endoscopy, Andover, MA 01810.

Recall: Z-1105-2007 · Initiated October 12, 2004

Recall

Recall Number
Z-1105-2007
Event Number
37614
Firm
Smith & Nephew, Inc Endoscopy Division
FEI Number
1643264
Product Code
DWQ
Status
Terminated
Root Cause
Other
Initiated
October 12, 2004
Posted
August 2, 2007
Terminated
August 2, 2007
Address
76 S. Meridian Ave, Oklahoma City, OK, 73107-6512

Description

Smith & Nephew TRIVEX System Pneumatic Footswitch, Part #7209791; distributed by Smith & Nephew, Inc. - Endoscopy, Andover, MA 01810.

Reason

Build-up of static charge in Pneumatic Footswitch tubing may cause control unit to reset to default settings in the operating room during surgery.

Action

Firm sent notification letters to consignees via overnight FedEx on 10/12-13/04. International customers were notified via letter on 11/10/04. Customers were asked to return footswitches for a replacement.

Distribution

Product sent to distributors nationwide and to the following countries: Australia, Austria, Canada, China, Germany, Italy, Japan, and Taiwan.

Quantity

274 units.