FDA Recall Open, Classified

Integra Miltex CryoSolutions Cartridges 10 Pack (23.5g N2O each), Model: C-CA-23, Item: 33516;

Recall: Z-1103-2025 · Initiated December 19, 2024

Recall

Recall Number
Z-1103-2025
Event Number
96067
Firm
New Medical Technologies Gmbh Wankdorffeldstrasse
FEI Number
3008252257
Product Code
GEH
Status
Open, Classified
Root Cause
Device Design
Initiated
December 19, 2024
Posted
February 5, 2025
Address
104 Bern Switzerland

Description

Integra Miltex CryoSolutions Cartridges 10 Pack (23.5g N2O each), Model: C-CA-23, Item: 33516;

Reason

During the assembly of the cartridge to its connecting body part, the valve may open prematurely resulting in partial or complete emptying of the cartridge before its intended use.

Action

On December 19, 2024 "Important Notice Regarding Your Device" letters were sent to customers. Corrective Actions: We have implemented the following corrective and preventive actions to address this issue: 1. Enhanced Testing: All cartridges are now being filled with a minimum specified volume (23.5g tolerance) and subjected to rigorous leakage testing. 2. Updated Specifications: We have revised the interface connection specifications to include precise tolerances and correlated these adjustments with the gas volume requirements. 3. Improved Instructions for Use (IFU): Additional safety precautions regarding gas leakage during assembly are now clearly highlighted in the Instructions for Use (IFU) provided with the product. We are committed to ensuring these actions eliminate the identified problem and prevent its recurrence. Our quality assurance team will also verify the effectiveness of these measures by closely monitoring product performance and feedback. What You Should Do: " Please ensure that all users review the updated Instructions for Use (IFU) to familiarize themselves with the additional safety measures, including warnings regarding the potential for premature gas leakage during cartridge connection. Aligned with that the gas cylinder shall be quickly screwed into the handpiece. " Please ensure that a new O-ring is assembled into the body part of the handpiece We sincerely apologize for any inconvenience this may cause and appreciate your understanding and cooperation. If you have any questions or require assistance, please do not hesitate to contact our customer service team. on January 31, 2025, URGENT: VOLUNTARY MEDICAL DEVICE CORRECTION letters were sent out to sub-accounts from their sole consignee. What You Should Do (Customers (i.e. Medical Facility, etc.)): 1. Please review the notice from UMP (Appendix 1). 2. Please ensure that you review the updated Instructions for Use (IFU) to familiarize yourselves with the additional safety measures, incl

Distribution

Worldwide - US Nationwide distribution in the states of AR, CA, CO, FL, GA, IL, IN, KY, LA, MA, MD, MN, MO, NC, NE, NJ, NV, NY, OH, PA, SC, TN, TX, VA, WA and the country of Canada.

Quantity

430 units (43 packs of 10)