SPiN Thoracic Navigation System Model: SYS-4000 (SYS-4000D)

Recall: Z-1103-2023 · Initiated December 13, 2022

Recall

Recall Number
Z-1103-2023
Event Number
91291
Firm
Olympus Corporation of the Americas
FEI Number
2429304
Product Code
JAK
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
December 13, 2022
Posted
February 7, 2023
Address
3500 Corporate Pkwy, Center Valley, PA, 18034-8229

Description

SPiN Thoracic Navigation System Model: SYS-4000 (SYS-4000D)

Reason

A grainy image is observed when using a SPiN Vision Single-Use Flexible Bronchoscope (models INS-7100 or INS-7130) with the SYS-4000 system due to an incorrect DVI cable supplied may result in prolonged procedure/surgery due to additional time to troubleshoot the device

Action

Olympus issued URGENT - Medical Device Field Corrective Action letter on 12/13/22. Letter states reason for recall, health risk and action to take: 1. Notify your users about this issue and stop using the single-use bronchoscopes (models INS-7100 or INS-7130) with SYS-4000 system until cable is replaced. Your users can continue to use the SYS-4000 using a reusable bronchoscope, with its respective video processor, without affecting their electromagnetic navigation (EMN) procedural experience. 2. Olympus requests that you acknowledge receipt of this letter by accessing the Olympus recall portal. Go to https://olympusamerica.com/recall. Enter the recall number 0418 Provide your contact information as indicated in the portal. An Olympus Field Service Representative will contact you to schedule a site visit to replace the DVI cable. 3. If you have distributed these devices outside your facility, please notify your customers of this matter immediately. Your notification to your customers may be enhanced by including a copy of this letter. Olympus requests that you report complaints, including patient injuries and adverse events, to Olympus. Please report complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1 to report complaints. Contact (647) 999-3203 or at [email protected] for any additional information or support.

Distribution

FL, GA, IN, MD, MS, NY, OH, PA, RI, SC, VA

Quantity

13 units