FDA Recall Terminated

Hitachi Altaire Magnetic Resonance Imaging System

Recall: Z-1103-06 · Initiated January 27, 2006

Recall

Recall Number
Z-1103-06
Event Number
34856
Firm
Hitachi Medical Systems America Inc
FEI Number
1528028
Product Code
LNH
Status
Terminated
Root Cause
Other
Initiated
January 27, 2006
Posted
June 10, 2006
Terminated
November 13, 2008
Address
1959 Summit Commerce Park, Twinsburg, OH, 44087-2371

Description

Hitachi Altaire Magnetic Resonance Imaging System

Reason

Circuit Failure-There is the potential risk of the Phase Loss Detector Safety Device (PHSEN PCB) malfunctioning in the event of a power loss. The circuit may fail to turn off the system main breaker when 2 of the 3 power phases are lost.

Action

The recalling firm sent a Device Correction Letter dated 1/27/06 to all of their customers. The letter informs the customers of the problem in the (PHSEN PCB) and notifies the customers that an HMSA Field Service Technician will be performing a PCB modification to the unit, on-site, during their next scheduled preventative maintenance visit. Should the customer wish to have the modification performed prior to the representative''s scheduled visit, the letter says that they can telephone their local Field Service Rep. and make arrangements to have the modification performed at their convenience.

Distribution

Nationwide-The device was distributed to consignees located throughout the United States.

Quantity

213 devices have been distributed and installed.