FDA Recall Terminated

V Series Monitor Mindray DS USA, Inc. 800 MacArthur Boulevard. Mahwah, NJ 07430 The Indications for Use for the V Series include the monitoring of the following human physiological parameters: ECG waveform derived from 3, 5, 6 and 12 lead measurements Heart Rate Pulse Oximetry (Sp02) ST Segment Analysis Arrhythmia Detection Non Invasive Blood Pressure (NIBP) Invasive Blood Pressure (IBP) Cardiac Output (CO) Respiratory Gasses Respiration Rate Temperature The V Series Monitor has the capability of performing IV Drug and Hemodynamic Calculations and interfacing with network devices

Recall: Z-1102-2012 · Initiated October 14, 2011

Recall

Recall Number
Z-1102-2012
Event Number
60976
Firm
Mindray DS USA, Inc. d.b.a. Mindray North America
FEI Number
2221819
Product Code
DRT
Status
Terminated
Root Cause
Process control
Initiated
October 14, 2011
Posted
February 24, 2012
Terminated
June 11, 2013
Address
800 MacArthur Blvd., Mahwah, NJ, 07430

Description

V Series Monitor Mindray DS USA, Inc. 800 MacArthur Boulevard. Mahwah, NJ 07430 The Indications for Use for the V Series include the monitoring of the following human physiological parameters: ECG waveform derived from 3, 5, 6 and 12 lead measurements Heart Rate Pulse Oximetry (Sp02) ST Segment Analysis Arrhythmia Detection Non Invasive Blood Pressure (NIBP) Invasive Blood Pressure (IBP) Cardiac Output (CO) Respiratory Gasses Respiration Rate Temperature The V Series Monitor has the capability of performing IV Drug and Hemodynamic Calculations and interfacing with network devices

Reason

A system database corruption may occur that causes the V Series Monitor to reset.

Action

The firm, Mindray, sent an "URGENT PRODUCT CORRECTIVE ACTION" letter dated October 14, 2011 on October 17, 2011 via certified mail, return receipt requested, signature required to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to contact their Mindray Service Representative at 1-800-288-2121, to arrange for software upgrade of the V Series Monitor(s) in their facility. If you have any questions, contact Manager, Regulatory Affairs and Clinical Affairs at 201-995-8407 or email: [email protected].

Distribution

Worldwide distribution: USA (nationwide) including states of: AL, CA, CO, IN, MO, NE, PR, and SD; and countries of: Australia, Chile, Columbia, United Kingdom, and Venezula.

Quantity

542 units US ; 153 units International