FDA Recall Terminated

MRT-1504/A4 - Titan X MRI System MRT-30 I0A/5 - Titan 3T MRI System Product Usage: MRI System.

Recall: Z-1101-2013 · Initiated October 15, 2012

Recall

Recall Number
Z-1101-2013
Event Number
64670
Firm
Toshiba American Medical Systems Inc
FEI Number
2020563
Product Code
LNH
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
October 15, 2012
Posted
April 11, 2013
Terminated
November 6, 2013
Address
2441 Michelle Dr P.O. Box 2068, Tustin, CA, 92780-7047

Description

MRT-1504/A4 - Titan X MRI System MRT-30 I0A/5 - Titan 3T MRI System Product Usage: MRI System.

Reason

It was found that there is a problem with the LCD monitor used in the console. The LCD monitor may become blank during use. Specifically, the backlight of the LCD monitor may turn off while the system is in use, causing the LCD monitor to become blank. However, the other parts of the system continue to operate normally, and the risk of a serious health hazard is therefore very low. The LCD Moni

Action

Toshiba sent an Urgent Medical Device Correction letter dated October 1, 2012 via USPS return receipt mail to all affected customers . The letter identified the affected product, problem and actions to be taken. The letter contains a return reply form that is to be faxed to TAMS for retention. For questions contact your local Toshiba representative at 800-521-1968.

Distribution

USA Nationwide Distribution including the states of : IL, NH, MA, MI, MO, PR, NV, NM, MN, ID, NY, OH, NC, AZ, MS, LA, OR, PA, CA, and CT.

Quantity

27 units