FDA Recall
Terminated
Synchron Systems Hemoglobin A1c (HbA1c) Reagent
Recall: Z-1101-04
·
Initiated July 30, 2004
Recall
- Recall Number
- Z-1101-04
- Event Number
- 28961
- Firm
- Beckman Coulter Inc
- FEI Number
- 2050012
- Product Code
- JJY
- Status
- Terminated
- Root Cause
- Other
- Initiated
- July 30, 2004
- Posted
- July 20, 2004
- Terminated
- August 30, 2004
- Address
- 200 S Kraemer Blvd, Brea, CA, 92822-6208
Description
Synchron Systems Hemoglobin A1c (HbA1c) Reagent
Reason
Results of tests are biased low due to a vendor change in formulation not brought to the attention of the recalling firm.