FDA Recall Terminated

AutoDELFIA Plate Processor , Fluorometer, for clinical use Product Number: 1235-5110 Product Usage: The Wallac 1235 AutoDELFIA automatic immunoassay system is designed to automatically perform assays using the DELFIA technology. DELFIA technology is based on the proven and widely used method of time resolved fluorometry

Recall: Z-1100-2017 · Initiated December 21, 2016

Recall

Recall Number
Z-1100-2017
Event Number
75949
Firm
PerkinElmer Health Sciences, Inc.
FEI Number
3006159931
Product Code
KHO
Status
Terminated
Root Cause
Software design
Initiated
December 21, 2016
Terminated
October 3, 2018
Address
940 Winter St, Waltham, MA, 02451-1457

Description

AutoDELFIA Plate Processor , Fluorometer, for clinical use Product Number: 1235-5110 Product Usage: The Wallac 1235 AutoDELFIA automatic immunoassay system is designed to automatically perform assays using the DELFIA technology. DELFIA technology is based on the proven and widely used method of time resolved fluorometry

Reason

AutoDELFIA System with Version 3.0 Software: Potential plate loading failures

Action

PerkinElmer issued letter on 12/21/16 to customers with AutoDELFIA Systems running version 3.0 Manager Software (revisons 1 and 2). The issue cannot be corrected by the customer therefore it is recommended the AutoDELFIA System be operated in the default English language variant until the AutoDELFIA Manager Software can be updated. The AutoDELFIA System with version 3.0 Manager Software performs as intended when the selected language variant is English. Customers are recommended to implement a software update to be deployed by PerkinElmer to resolve the issue. AutoDELFIA Manager Software version 3.0 revision 1 or 2 will be updated through the correction to revision 3 or the latest released revision

Distribution

Worldwide Distribution - US Nationwide in the state of MN and countries of China, India, Philippines, France, South Korea, Vietnam, Taiwan, Morocco, Kazachstan, UK, Spain, and Israel,

Quantity

1 unit (US) 6 (OUS)